Labconco Acid-Resistant Centrivap Vacuum Concentrator System w/ -50C Cold Trap
The Labconco Acid-Resistant Centrivap Vacuum Concentrator System w/ -50C Cold Trap is designed to rapidly concentrate small samples containing corrosive chemicals, including 0.1 to 1% TFA, hydrochloric acid, acetic acid, hydrobromic acid, ammonium hydroxide, and sodium hydroxide. Utilizing centrifugal force, vacuum, and controlled heat, it ensures efficient concentration while components in the vapor path are constructed from corrosion-resistant materials for durability. The exterior features brushed stainless steel and glacier white, epoxy-coated steel with blue accents, complemented by a glass lid and phenol-free lid gasket. A PTFE-coated aluminum chamber with a Hastelloy shaft enhances resistance to aggressive chemicals.
This concentrator includes a 300-watt heater, microprocessor-controlled programming for precise settings (1 to 999 minutes for run and heat time, and temperature adjustable from OFF to HI at 1°C increments), and a memory that stores up to 9 programs. Its user-friendly control panel has RUN/STOP, PREHEAT, and set point SELECT buttons, plus three Quick-Start buttons for convenience. An LCD displays program information, temperature, and time remaining. The brushless motor achieves speeds up to 1,725 RPM, while the Quick-Stop System halts operation safely and efficiently. Additional features include an automatic vacuum release valve, built-in vacuum delay, and audible alarm signaling completion. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Labconco Acid-Resistant Centrivap Vacuum Concentrator System w/ -50C Cold Trap includes:
- > Acid-Resistant CentriVap Concentrator
- > Clear Canister
- > Acid Insert
- > 7450701 PTFE-Coated Fixed Angle Rotor
- > CentriVap Cold Trap -50C
- > Ilmvac MPC 601 TP Diaphragm Vacuum Pump
- > Power Cords
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Labconco Acid-Resistant Centrivap Vacuum Concentrator System w/ -50C Cold Trap Specifications
Labconco Acid-Resistant Centrivap Vacuum Concentrator
|
Maximum Speed
|
1,725 RPM
|
Temperature Range
|
Up to 100°C
|
Timer
|
1 to 999 minutes
|
Ultimate Vacuum
|
2 mbar with included external pump
|
Rotor Capacity
|
132 x 1.5/2.0 ml
|
Dimensions (W x D x H)
|
14.7 x 19.3 x 11.7 in. (36.3 x 55.1 x 42.2 cm)
|
Weight
|
65 lb. (29.0 kg)
|
Electrical Requirements
|
115V, 60Hz, 12A
|
Labconco CentriVap Cold Trap -50C
|
Cold Trap Temperature
|
-50°C
|
Collector Volume
|
4L
|
Type / Application
|
Aqueous
|
Lid Type
|
Acrylic
|
CentriVap Optional Accessories
|
Clear canisters and chemical trap inserts, glass trap, two-port manifold
|
Material
|
Brushed stainless steel and powder-coated steel
|
Dimensions (W x D x H)
|
14.5 x 24.2 x 13.8 in. (36.8 x 61.4 x 35 cm)
|
Weight
|
80 lb. (36 kg)
|
Electrical Requirements
|
115V, 60Hz, 5.4A
|
Ilmvac MPC 601 TP Diaphragm Vacuum Pump
|
Ultimate Vacuum
|
2 mbar
|
Displacement
|
2.9 CFM at 60Hz
|
Noise Level
|
45 dB(A)
|
Weight
|
40.3 lb. (18.3 kg)
|
Electrical Requirements
|
115V, 50/60Hz
|
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|