IKA MTS 2/4 D S1 Microplate Shaker
The IKA MTS 2/4 D S1 Microplate Shaker is tailored for efficient agitation of two or four microtiter plates simultaneously. It features an adjustable speed range from 0 to 1100 RPM. This shaker employs gentle orbital motion with a 4 mm diameter to ensure thorough mixing without compromising sample integrity. Compact and robust, it is an ideal tool for laboratory environments requiring reliable plate shaking capabilities.
All of our equipment is fully tested before shipping.
The IKA MTS 2/4 D S1 Microplate Shaker includes:
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
IKA MTS 2/4 D S1 Microplate Shaker Specifications
Type of movement |
Orbital |
Shaking stroke |
4 mm |
Max load (incl. attachment) |
2 kg |
Input power |
19 W |
Output power |
5 W |
Speed range |
0 - 1100 rpm (infinitely adjustable) |
Speed display |
Scale 0 - 1100 |
Dimensions (W x H x D) |
157 x 247 x 130 mm (without attachment) |
Weight |
5,7 kg |
Voltage |
115 VAC / 230 VAC |
Frequency |
50/60 Hz |
Click Here to View all the Technical Specs for the IKA MTS 2/4
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|