Hettich Rotanta 460R Refrigerated Centrifuge, Cat # 5605-01
The Hettich Rotanta 460R Refrigerated Centrifuge, Cat # 5605-01 reaches maximum speeds up to 15,000 RPM with a maximum RCF up to 24,400 x g and a temperature range from - 20 to +40 °C . Timer up to 99 minutes, featuring 3 programmable run memories. Safety features include powered lid locking, lid dropping protection, lid lock and hold, automatic rotor recognition, motor overheating protection, chamber overheating protection, and imbalance switch-off. This rotor reaches maximum 4,600 RPM. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Hettich Rotanta 460R Refrigerated Centrifuge includes:
- > 5624 Swing-Out Rotor
- > 5628 Buckets w/ Microplate Adapters
- > Calibration Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Hettich Rotanta 460R Refrigerated Centrifuge, Cat # 5605-01 Specifications
Maximum Speed
|
15,000 RPM
|
Maximum RCF
|
24,400 x g
|
Timer
|
1 sec – 99 min : 59 sec, continuous run, short cycle
mode (impulse key)
|
Maximum Capacity
|
4 x 750 ml, swing - out rotor
|
Dimensions (H x W x D)
|
453 x 770 x 707 mm
|
Temperature Range
|
from -20°C to 40°C
|
Electrical Requirements
|
100-127V, 60 Hz, 17A
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|