Heidolph Titramax 100 Vibrating Platform Shaker
The Heidolph Titramax 100 is a compact and space-saving unit that accepts load capacities of 2 kg and has space available for 4 microtiter plates. The 1.5-mm vibration orbit performs gentle mixing while the analog process timer allows for unattended operation with a 1 to 120 minute timer with speeds from 150 to 1,350 RPM.
The Heidolph Titramax 100 Vibrating Platform Shaker includes:
- > 220 x 200 mm Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Heidolph Titramax 100 Vibrating Platform Shaker Specifications
Speed Range
|
150 to 1350 RPM
|
Timer
|
1 to 120 minutes
|
Maximum Load
|
2 kg
|
Ambient Operating
Temperature
|
Up to 65°C
|
Dimensions (W x D x
H)
|
245 x 310 x 125 mm
|
Electrical
Requirements
|
115V, 60Hz
|
Platform Dimensions
|
220 x 220 mm
|
Orbit/Stroke
|
1.5 mm
|
Motion
|
Vibrating
|
Click Here to View all the Technical Specs for the Titramax 100
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|