Heidolph Hei-VAP Precision ML/G5 Rotary Evaporator System with Chiller
The Heidolph Hei-VAP Precision ML/G5 Rotary Evaporator System with Chiller is designed for the most demanding applications, requiring the finest integrated vacuum control and cooling capabilities. The Hei-VAP series of rotary evaporator systems all feature a stainless steel heating bath with 2 independent safety circuits, which can accommodate up to a 5 liter flask at a maximum temperature of 210°C. This configuration is equipped with the Motor Lift, allowing for precise height adjustment of the evaporating flask at a rather quick adjustment speed of 30 mm/second. With the Hei-VAP Precision Controller, the user has access to an extensive array of user-defined parameters on an easy-to-use 4.3" LCD display, making this rotary evaporator perfect for reproducible applications and fully automated processes in the laboratory. The Heidolph Hei-VAP Precision also features the use of a vapor temperature sensor, which displays the actual temperature right on the Hei-VAP Precision controller. (The AUTOAccurate sensor however is not included, and is only compatible with the G3 and G6 Glassware types). Also included with this system is the Brand-New Polyscience LM61GX1A110C LM6 Chiller, 560W, which provides reliable cooling from -10°C to 30°C. This chiller also features a cooling capacity of 560W and a maximum flow rate of 7.9L/min (2.1 gpm).
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Heidolph Hei-VAP Precision ML/G5 Rotary Evaporator System with Chiller includes:
- > Rotavac Valve Tec Diaphragm Vacuum Pump
- > Heidolph Hei-VAP Precision Controller
- > Heidolph Hei-VAP Heating Bath
- > Brand-New Polyscience LM61GX1A110C LM6 Chiller, 560W
- > Coil Glassware (Dry Ice Condenser is Pictured)
- > 1 Liter Receiving Flask
- > 1 Liter Evaporating Flask
- > Vacuum Valve w/ Woulff Bottle
- > Vacuum Box w/ Interface Cable
- > Vacuum Sensor
- > Vapor Temperature Sensor
- > All Necessary Power Cords, Tubing, and Connectors
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Heidolph Hei-VAP Precision ML/G5 Rotary Evaporator System with Chiller Specifications
Heidolph Hei-VAP Precision ML/G5 Rotary Evaporator
|
Maximum Flask Size
|
5L
|
Hand / Motor Lift
|
Motor Lift
|
Motor Lift Adjustment Speed
|
30 mm/s
|
Motor Lift Height Adjustment
|
155 mm
|
Rotation Speed Range
|
10 to 280 RPM
|
Rotation Speed Control
|
Digital
|
Drive
|
Brushless DC Motor with Electronic Speed Control
|
Electrical Requirements
|
115V, 50/60Hz, 1400W
|
Heidolph Hei-VAP Heating Bath
|
Heating Capacity
|
1,300W
|
Temperature Range
|
20°C to 210°C
|
Temperature Accuracy
|
± 1°C
|
Overheat Protection
|
Cut off at 5°C over set temperature
|
Temperature Control
|
Digital
|
Secondary Overtemp Cutoff
|
250°C
|
Material
|
Stainless Steel AISI 316L
|
Bath Volume
|
4 to 5L
|
Heidolph Rotavac Valve Tec Diaphragm Vacuum Pump
|
Suction Capacity
|
0.9 m^3/h
|
Ultimate Vacuum
|
12 mbar
|
Diaphragm Stage
|
Two-stage
|
Maximum Motor Speed
|
1,800 RPM
|
Weight
|
13.2 lb. (6 kg)
|
Dimensions without Condenser
|
7.7 x 9.6 x 12.2 in (19.5 x 24.5 x 31 cm)
|
Polyscience LM61GX1A110C LM6 Chiller Specifications
|
Temperature Range
|
-10°C to 30°C
|
Temperature Stability
|
± 0.1°C
|
Cooling Capacity
|
560W
|
Maximum Flow Rate
|
7.9L/min (2.1 gpm)
|
Electrical Requirements
|
120V, 60Hz, 12A, 1Ph
|
- Inspect all glassware for chips or cracks, reapplying vacuum grease to stopcocks to ensure a secure seal
- Examine displays, replacing any dim or dead pixels to maintain clear visibility
- Verify drive motor RPM using a tachometer for accurate rotational speed
- Confirm the ease of raising and lowering the tower, lubricating as needed for smooth operation
- Check the Woulff bottle for any leaks or structural breakages, ensuring airtight performance
- Disassemble and inspect the vacuum pump, cleaning and replacing seals or components as necessary
- Inspect internal wiring for each component, checking for any frays, cuts, or potential breakages
- Examine all hoses, hose barbs, and seals for signs of wear or leaks, ensuring integrity
- Inspect the vapor duct seal and replace if compromised to maintain vacuum efficiency
- Verify and calibrate the vacuum controller for accurate pressure regulation
- Confirm all hoses and cables are securely fastened to prevent disconnections or leaks
- Check heat bath temperatures against settings to ensure precise thermal control
- Confirm the final system vacuum level meets manufacturer specifications
- Ensure cosmetic condition meets standards, with no damage impacting function or appearance
Technician Inspection and Recertification Checklist: Rotary Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect all glassware for chips or cracks, apply vacuum grease to stopcocks
|
Completed
|
|
Examine displays and replace dim/dead pixels
|
Completed
|
|
Verify drive motor RPM with tachometer
|
Completed
|
|
Confirm ease of raising/lowering tower, lubricate if needed
|
Completed
|
|
Check Woulff bottle for leaks or structural issues
|
Completed
|
|
Disassemble and inspect vacuum pump, clean/replace parts as needed
|
Completed
|
|
Inspect internal wiring for frays, cuts, or breakages
|
Completed
|
|
Examine hoses, hose barbs, and seals for leaks
|
Completed
|
|
Inspect vapor duct seal, replace if compromised
|
Completed
|
|
Verify and calibrate vacuum controller
|
Completed
|
|
Confirm hoses and cables are securely fastened
|
Completed
|
|
Check heat bath temperatures for accurate control
|
Completed
|
|
Confirm final system vacuum level meets specifications
|
Completed
|
|
Ensure cosmetic condition meets standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|