Hydraulic System
|
Dual piston in series pump with proprietary servo-controlled variable
stroke drive, floating pistons and active inlet valve
|
Setable Flow Range
|
0.001 — 10 ml/min, in 0.001 ml/min increments
|
Flow Range
|
0.2 — 10.0 ml/min
|
Flow Precision
|
<0.07% RSD, or <0.02 min SD whatever is greater, based on
retention time at constant room temperature
|
Flow Accuracy
|
±1% or 10µl/min whatever is greater
|
Pressure
|
Operating range 0 — 40 MPa (0 — 400 bar, 0 — 5880 psi) up to 5 ml/min
Operating range 0 — 20 MPa (0 — 200 bar, 0 — 2950 psi) up to 10 ml/min
|
Pressure Pulsation
|
< 2 %amplitude (typically < 1 %), at 1 ml/min isopropanol, at
all pressures > 10 bar (147 psi)
|
Compressibility Compensation
|
User-selectable, based on mobile phase compressibility
|
Recommended PH Range
|
1.0 — 12.5, solvents with pH < 2.3 should not contain acids which
attack stainless steel
|
Control and Data Evaluation
|
Agilent ChemStation for LC
|
Analog Output
|
For pressure monitoring, 2 mV/bar, one output
|
Communications
|
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN optional
|