Gilson PIPETMAN Neo Pipette Set
- Neo Pipette Set: PVDF handle, stainless steel ejector.
- Volumes: 0.2µL to 10mL, enhanced piston stroke smoothness.
- Includes: 7 Pipettes (P2N to 2 x P1000N's), stand, 1-Year Warranty.
- Specifications: Adjustable Volume, Manual Air Displacement.
Go to Product Details
No Lead Time and Free Shipping!
Gilson P2, P10, P20, P100, P200, P1000, P5000 Pipette Set
This eight-piece Gilson Pipette Set includes one each of the P2, P10, P20, P100, P200, P1000, and P5000 pipettes and a Gilson pipette stand. The Gilson PIPETMAN set is a variable volume air displacement pipette set for dispensing volumes up to 5000ul with precision accuracy and ease. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Gilson P2, P10, P20, P100, P200, P1000, P5000 Pipette Set includes:
- > Calibration
- > Pipette Stand
- > Full 180 Day Parts and Labor Warranty
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Thoroughly inspect and clean all internal components; replace damaged parts
|
Completed
|
|
Replace essential wear components like O-rings, seals, and friction rings
|
Completed
|
|
Examine plunger and seal holders for wear or damage; replace if needed
|
Completed
|
|
Verify no leakage is present by aspirating sample and checking for fluid loss
|
Completed
|
|
Confirm Mettler Toledo AB104-S balance calibration for accurate measurements
|
Completed
|
|
Perform detailed gravimetric calibration and record findings in report
|
Completed
|
|
Clean all external parts meticulously to remove any contaminants
|
Completed
|
|
Test tip ejector for smooth, reliable operation; correct misalignments
|
Completed
|
|
Affix dated calibration sticker to indicate verified calibration
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|