Gilson P10, P20, P100, P200, P1000, Pipette Set Classic Large Plunger
Volume Range 1 to 1000 uL
Pipette Type Manual Air Displacement
- Continuously adjustable air-displacement pipettes known for legendary accuracy and precision
- Includes five single-channel models covering volumes from 1 to 1000 uL
- Constructed with highly resistant materials like PVDF and stainless steel for durability and reliability
- Economical solution with a long lifespan, ideal for routine laboratory use
- Easy to maintain and service with user-accessible parts
- Features a fully autoclavable tip holder and ejector for sterility and cleanliness
The Gilson PIPETMAN Classic series is one of the world standards in pipetting and offers a wide range of continuously adjustable, air-displacement pipettes with legendary accuracy and precision. This pipette set includes five single channel models, covering a range of volumes from 1 uL to 1000 uL. Made of highly resistant materials like PVDF and stainless steel, the PIPETMAN Classic series offers uncontested durability and reliability. Built-to-last, it is the economical solution for years of pipetting. Routine cleaning and maintenance are all that is required to keep PIPETMAN Classic series pipettes in top condition for years. The PIPETMAN Classic series design allows for easy access to user-serviceable parts, along with a fully autoclavable tip holder and ejector.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Gilson P10, P20, P100, P200, P1000 Pipette Set Classic Large Plunger includes:
- > Gravimetric Analysis Calibration Reports
- > Full 180 Day Parts and Labor Warranty
Pipette Model
|
Volume (µL)
|
Maximum Systematic Error (µL)
|
Maximum Random Error (µL)
|
P10
|
1
5
10
|
± 0.12
± 0.12
± 0.12
|
≤ 0.08
≤ 0.08
≤ 0.08
|
P20
|
2
5
10
20
|
± 0.2
± 0.2
± 0.2
± 0.2
|
≤ 0.1
≤ 0.1
≤ 0.1
≤ 0.1
|
P100
|
10
50
100
|
± 0.8
± 0.8
± 0.8
|
≤ 0.3
≤ 0.3
≤ 0.3
|
P200
|
20
100
200
|
± 1.6
± 1.6
± 1.6
|
≤ 0.6
≤ 0.6
≤ 0.6
|
P1000
|
100
500
1000
|
± 8.0
± 8.0
± 8.0
|
≤ 3.0
≤ 3.0
≤ 3.0
|
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Thoroughly inspect and clean all internal components; replace damaged parts
|
Completed
|
|
Replace essential wear components like O-rings, seals, and friction rings
|
Completed
|
|
Examine plunger and seal holders for wear or damage; replace if needed
|
Completed
|
|
Verify no leakage is present by aspirating sample and checking for fluid loss
|
Completed
|
|
Confirm Mettler Toledo AB104-S balance calibration for accurate measurements
|
Completed
|
|
Perform detailed gravimetric calibration and record findings in report
|
Completed
|
|
Clean all external parts meticulously to remove any contaminants
|
Completed
|
|
Test tip ejector for smooth, reliable operation; correct misalignments
|
Completed
|
|
Affix dated calibration sticker to indicate verified calibration
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|