GE Biacore 3000 SPR Biomolecular Interaction Analyzer
The GE Biacore 3000 SPR Biomolecular Interaction Analyzer is an established, label-free system for detailed studies of biomolecular reactions. The Biacore 3000 delivers comprehensive characterization of the interaction, answers questions about the rate constants, affinity, specificity, and determines the active concentration of components. The unique ability to recover and transfer interaction and further characterization makes the system highly applicable to fast identification of unknown interactants. This analyzer is suitable for studying the interactions of proteins, protein conjugates, nucleic acids, lipid micelles, and even larger particles such as viruses and whole cells. Other major applications include measurement of kinetic constants, detection of binding partners and patterns in screening applications, monitoring of enzymatic processes, defining potential drug targets, and more.
The GE Biacore 3000 SPR Biomolecular Interaction Analyzer is designed for real-time biomolecular interaction analysis using surface plasmon resonance (SPR) technology. This technique monitors the formation and dissociation of biomolecular complexes on a sensor surface as the interaction occurs. By covalently attaching one molecule (the ligand) to the surface, the binding of another molecule in a solution (the analyte) with the ligand is followed. Measurements are made under conditions of continuous flow, where the sensor surface forms one wall of the flow cell. The Biacore 3000 measures surface protein concentrations from a few pg/mm^2, and can monitor changes in the surface concentration on a time scale down to 0.1 seconds. This technology allows for determination of binding specificity, kinetics, and concentration with analytes ranging in size from about 150 to 10^6 g/mole (Daltons) or more. Operation of the Biacore 3000 can be preprogrammed for analysis of up to 192 samples ( 2 x 96-well microplates).
The GE Biacore 3000 SPR Biomolecular Interaction Analyzer includes:
- > Windows Data Station
- > Biacore 3000 Control Software Installed
- > Autosampler Rack
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
GE Biacore 3000 SPR Biomolecular Analyzer Specifications
Detection Unit
|
Specification
|
Light Source
|
LED
|
Wavelength
|
760 nm
|
Band Width
|
< 5 nm
|
Refractive Index Range
|
1.33 - 1.40 (0 - 70,000 RU)
|
Baseline Drift
|
Typically < ± 0.3 RU/minute
|
Data Output
One Flow Cell
Two Flow Cells
Four Flow Cells
|
0.1, 1 or 10 measurements/s (0.1 - 10 Hz)
0.1, 1 or 5 measurements/s (0.1 - 2 Hz)
0.1, 1 or 2 measurements/s (0.1 - 5 Hz)
|
Noise
|
Typically < 0.3 RU RMS (10 Hz)
< 0.1 RU RMS (1 Hz)
|
Temperature Range
|
4°C to 40°C (not lower than 20°C below ambient temperature)
|
Temperature Stability
|
Better than ± 3 x 10^-3°C/min
|
Temperature Equilibration Time
After 5°C Step
After 10°C Step
|
Typically < 60 minutes
Typically < 120 minutes
|
Molecular Weight Detection
|
> 180 Da (Daltons)
|
Pump
|
Specification
|
Flow Rate Range
|
1 - 100 µl/minute user-controlled
|
Increment
|
Steps of 1 µl/minute
|
Flow Rate Accuracy
|
< ± 2%
|
Flow Rate Precision
|
< ± 1%
|
Pressure Range
|
< 0.2 MPa
|
Autosampler
|
Specification
|
Dispensing Volume Range
|
5 - 500 µl (including air segment)
|
Increment
|
Steps of 1 µl
|
Accuracy
|
< 1% deviation for volumes greater than or equal to 50 µl
|
Precision
50 µl
5 µl
|
< 0.4% CV
< 2% CV
|
Carry Over Between Vials
|
Typically < 1 µl without septum
Typically < 0.1 µl with septum
|
Sample Capacity per Rack
Microplate
Rack Type A
Rack Type B
Rack Type C
|
96-well (0.2 ml/well)
5 Vials, diam. 16 mm (4 ml) 12 Vials, diam. 9 mm (1.5 ml) 40 Vials, diam. 7 mm (0.7 ml)
60 Vials, diam. 9 mm (1.5 ml)
24 Vials, diam. 10.8 mm (Eppendorf 1.5/2.0 ml)
|
Maximum Vial Height
|
55 mm
|
Integrated µ-Fluidic Cartridge
|
Specification
|
Number of Flow Cells
|
4
|
Valves
|
Pneumatic Membrane Valves
|
Injection Volume to Flow Cell
|
5 - 750 µl, steps of 1 µl
|
Carry Over from Previous Injection
|
< 0.1%
|
Sample Loop Capacity
|
120 µl (maximum injection volume of 100 µl)
|
General
|
Specification
|
Ambient Temperature Range
|
15°C to 35°C
|
Ambient Humidity
|
20 to 85% RH
|
Dimensions (L x W x H)
|
29.9 x 13.8 x 24 in (760 x 350 x 610 mm)
|
Weight
|
110 lbs (50 kg)
|
Electrical Requirements
|
100-120V / 220-240V, 47-63 Hz, 580 VA
|
GE Biacore 3000 SPR Biomolecular Interaction Analyzer Manual
Click Here to View the Technical Specs for the GE Biacore 3000 SPR Analyzer
- Verify power supply functionality to ensure stable operation of all components
- Confirm precise seating and alignment of the electrode cartridge
- Detach and perform ultrasonic cleaning on electrode pin assembly
- Validate communication between analyzer and software for seamless integration
- Ensure Agilent lid opens and closes smoothly without abrasion against the casing
- Inspect internal wiring for any shorts, tears, or compromised insulation
- Check autofocus motors for full movement range and operational accuracy
- Execute comprehensive autofocus system diagnostic analysis
- Perform detailed diagnostics for electrode and diode functionality
- Complete remaining 2100 Expert diagnostic tests to verify optimal functionality
- Inspect power cords and interface cables for wear, tears, or exposed wiring
- Verify software is fully operational and compatible with the updated computer system
Technician Inspection and Recertification Checklist: Bioanalyzers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply functionality
|
Completed
|
|
Confirm electrode cartridge seating and alignment
|
Completed
|
|
Clean electrode pin assembly using ultrasonic cleaning
|
Completed
|
|
Validate communication with software
|
Completed
|
|
Confirm lid opens and closes without abrasion
|
Completed
|
|
Inspect internal wiring for damage or shorts
|
Completed
|
|
Check autofocus motor functionality
|
Completed
|
|
Run autofocus diagnostic analysis
|
Completed
|
|
Perform electrode and diode diagnostics
|
Completed
|
|
Complete all 2100 Expert diagnostic tests
|
Completed
|
|
Inspect power and interface cables for damage
|
Completed
|
|
Confirm software and computer compatibility
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|