Fisher Vortex Genie 2 G560 Vortexer
Fisher Vortex Genie 2 G560 Vortexer features variable Manual Speed Controls, Touch Function and Automatic Function, 3? Platform Head. The Fisher Vortex Genie 2 G560 offers a speed range of 600 to 3200 rpm. All of our equipment is fully tested before being shipped.
The Fisher Vortex Genie 2 G560 Vortexer includes:
- > Rubber Dimpled Head
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Fisher Vortex Genie 2 G560 Vortexer
Environmental
|
0C
- 38C
|
Electrical Requirements
|
120 V, 60 Hz, 0.65 amps
|
Weight
|
4 kg, 8.8 lbs
|
Base Dimensions (D x W x H)
|
6.5 x 4.8 x 6.5 in
|
Speed
|
600-3200 RPM
|
- Verify power supply and PCB output correct, stable voltages across all required circuits
- Confirm the correct platform head is installed according to specifications
- Inspect the display for dim or dead pixels, replacing any compromised sections to maintain clear visibility
- Confirm full functionality of all buttons and knobs, ensuring accurate input response
- Verify that touch, pulse, and continuous modes operate fully and as intended
- Inspect all internal wiring for breakages, tears, or wear, replacing any underperforming components to ensure reliability
- Disassemble and thoroughly clean the internal casing to remove any accumulated debris
- Check that the feet are intact and confirm the vortexer is properly balanced on a flat surface
- Conduct a comprehensive cosmetic inspection to ensure all exterior surfaces meet quality standards
Technician Inspection and Recertification Checklist: Vortexer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply and PCB output stable, correct voltages across circuits
|
Completed
|
|
Confirm correct platform head is installed according to specifications
|
Completed
|
|
Inspect display for dim or dead pixels; replace compromised sections for clarity
|
Completed
|
|
Confirm functionality of all buttons and knobs for accurate input response
|
Completed
|
|
Verify full operation of touch, pulse, and continuous modes as intended
|
Completed
|
|
Inspect internal wiring for damage; replace underperforming components for reliability
|
Completed
|
|
Disassemble and clean internal casing to remove accumulated debris
|
Completed
|
|
Check feet for integrity and confirm vortexer is balanced on flat surface
|
Completed
|
|
Conduct comprehensive cosmetic inspection to ensure quality of exterior surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|