Fisher Scientific Isotemp Dual Digital Water Bath Model 2322
Temperature Range Ambient to 100°C
Temperature Uniformity ±0.24°C at 37°C
Capacity 5L and 10L (1.3 gal and 2.6 gal)
- Suitable for a wide range of clinical procedures, including incubation, inactivation, and agglutination. Also ideal for serological, pharmaceutical, biomedical, and industrial applications.
- Designed for use with either water and bath oil mediums.
- The dual chamber design features independent controls for each chamber.
- Microprocessor control ensures accurate and reliable temperature settings.
- Dual thermostats provide precision as one thermostat controls the set point temperature while the other acts as a safety backup to prevent overheating
- Automatic self-test of vital circuitry each time the unit is turned on ensures smooth operation.
The Fisher Scientific Isotemp Dual Digital Water Bath Model 2322 features dual chambers, 5L capacity and 10L. This water bath has a temperature range from ambient to 100°C with a temperature uniformity of ±0.24°C at 37°C. Isotemp Water Baths offer the versatility needed to handle virtually any clinical laboratory procedure, including: incubation, inactivation, agglutination, serological, pharmaceutical, biomedical, and industrial procedures.
All of our equipment is fully tested to perform at factory specifications before being shipped.
Fisher Scientific Isotemp Dual Digital Water Bath Model 2322 includes:
- Covers
- Calibration
- Power Cord
- Operator's Manual
- Full 180 Days Parts and Labor Warranty
Fisher Scientific Isotemp Dual Digital Water Bath Model 2322 Specifications
Temperature Range
|
Ambient to 100°C
|
Temperature Uniformity
|
±0.24°C at 37°C
|
Control
|
Digital
|
Capacity
|
5L and 10L (1.3 gal and 2.6 gal)
|
Bath Dimensions (L x W x H)
|
5L Bath: 6.1 x 11.8 x 5.9 in (15.4 x
30 x 15 cm)
10L Bath: 11.9 x 13 x 5.9 in (30.1 x
33 x 15 cm)
|
Electrical Requirements
|
120V, 50/60Hz, 9A
|
Temperature Stability
|
±0.5°C at 37°C
|
- Confirm power supply and PCB are outputting stable, correct voltages across all essential circuits to ensure reliable operation
- Inspect the display for any dim, flickering, or dead pixels, replacing any compromised sections to maintain full clarity and accurate readouts
- Verify that all buttons, knobs, and switches are functioning properly, providing precise and responsive control across all settings
- Conduct a thorough inspection of the water reservoir for any holes, rust, or other structural damage that may impact its integrity
- Fill the reservoir with water, checking closely for any external or internal leaks that could affect performance
- Examine the lid for any signs of wear or damage, replacing hinges and handles if necessary to ensure a secure and proper seal
- Inspect for any internal leaks originating from the water reservoir, confirming the unit’s internal plumbing integrity
- Conduct a comprehensive inspection of internal wiring for any breakages, tears, or wear; replace any underperforming or damaged components to maintain safe operation
- Examine internal insulation for signs of burns, breaks, or deterioration, replacing as needed to ensure thermal efficiency and safety
- Confirm that all feet are intact and that the bath sits level and balanced on a flat surface to prevent movement or instability
- Calibrate both temperature accuracy and uniformity using a NIST-traceable temperature probe, verifying compliance with specification standards
- Perform a detailed cosmetic inspection to confirm that all external surfaces and finishes meet aesthetic and quality standards
Technician Inspection and Recertification Checklist: Water Bath
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm power supply and PCB output stable, correct voltages across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace for clarity
|
Completed
|
|
Verify functionality of all buttons, knobs, and switches for responsive control
|
Completed
|
|
Inspect water reservoir for holes, rust, or structural damage
|
Completed
|
|
Fill reservoir with water; check for external or internal leaks
|
Completed
|
|
Examine lid for wear or damage; replace hinges or handles as needed
|
Completed
|
|
Inspect for internal leaks from water reservoir to confirm plumbing integrity
|
Completed
|
|
Check internal wiring for damage; replace underperforming components for safety
|
Completed
|
|
Examine internal insulation for burns, breaks, or deterioration; replace as needed
|
Completed
|
|
Confirm feet are intact and bath is balanced on flat surface
|
Completed
|
|
Calibrate temperature accuracy and uniformity with NIST-traceable probe
|
Completed
|
|
Perform cosmetic inspection to confirm quality of external surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|