Popular Alternative Choice
IKA Oven 125 Basic Drying Oven
- For tempering, drying, aging, heating in labs.
- 125L, RT + 5°C to 250°C, USB controllable.
- Includes: trays, tilt protection, USB cable.
- 1500W, 0.1K resolution, 6 max trays, 74kg.
Go to Product Details
Brand New!
Fisher Scientific Isotemp 650G Gravity Oven
The Fisher Scientific Isotemp 650G Gravity Oven is in excellent condition. The Isotemp 650G delivers excellent temperature uniformity and stability. The 650G features 4 minute temperature recovery, timer, auto calibration routine for stable temperatures over time, and overtemperature alarm. Corrosion resistant stainless steel interior and doors that open past 180° for easy access. Applications include sample drying, dry sterilization, processing electronics, aging tests, etc. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Fisher Scientific Isotemp 650G Oven includes:
- > 2 Shelves
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific Isotemp 650G Gravity Oven Specifications
Capacity
|
5 cu ft
|
Temperature Range
|
50° to 225°C
|
Control Sensitivity
|
±0.5°C
|
Temperature Uniformity
|
±4.5°C at 200°C
|
Resolution
|
1°C
|
Recovery Time at 200°C
|
4 minutes
|
Rise Time to 225°C
|
120 minutes
|
Air Exchanges
|
12 per hour
|
BTU/hour Output
|
1150 at 100°C
2040 at 200°C
|
Interior Dimensions (W x D x H)
|
18 x 18 x 26.5 in
|
Electrical Requirements
|
120V, 60Hz, 11A, 1300W
|
Fisher Scientific Isotemp 650G Manual
Click Here to View all the Technical Specs for the Isotemp 650G
- Verify power supply and PCB outputs are within specified voltage ranges
- Inspect display for any dim or dead pixels, replacing as needed to ensure visibility
- Confirm full functionality and responsiveness of all control buttons
- Examine all internal wiring for frays, cuts, or damage, ensuring secure and reliable connections
- Inspect door seals and gaskets for integrity, replacing any compromised components
- Evaluate door handle and locking mechanism for wear or issues, replacing if performance is impaired
- Inspect and clean internal fan; replace if performance does not meet requirements
- Calibrate set temperature using a NIST-traceable probe for precise accuracy
- Confirm functionality of over-temperature alarm and indicator light
- Verify internal outlet functionality, if applicable, to ensure it meets operational standards
- Inspect internal lighting for full functionality and brightness, if applicable
- Perform a cosmetic inspection to confirm the oven’s exterior and interior meet quality standards
Technician Inspection and Recertification Checklist: Ovens
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply and PCB outputs are within specified voltage ranges
|
Completed
|
|
Inspect display for dim or dead pixels; replace if needed
|
Completed
|
|
Confirm full functionality of all control buttons
|
Completed
|
|
Examine internal wiring for damage; ensure secure connections
|
Completed
|
|
Inspect door seals and gaskets, replacing any compromised components
|
Completed
|
|
Evaluate door handle and locking mechanism, replacing if worn
|
Completed
|
|
Inspect and clean internal fan; replace if performance is insufficient
|
Completed
|
|
Calibrate set temperature using a NIST-traceable probe
|
Completed
|
|
Confirm over-temperature alarm and indicator light functionality
|
Completed
|
|
Verify internal outlet functionality, if applicable
|
Completed
|
|
Inspect internal lighting for full functionality and brightness
|
Completed
|
|
Perform cosmetic inspection to confirm quality standards of exterior and interior
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|