|
|
|
|
Popular Alternative Choice
Polyscience WBE02A11B 2L Digital Water Bath (Ambient +5 to 99C), 120V, 60Hz
- Polyscience WBE02A11B 2L Digital Water Bath: Ideal for various lab applications.
- Displays actual & set temperature; integrated timer.
- Three push-button pre-sets for frequently used temperatures.
- 304 Stainless Steel reservoir; safety thermostats; 2-year warranty.
Go to Product Details
Free Shipping!
Fisher Scientific Isotemp 102 Water Bath
The Fisher Scientific Isotemp 102 Water Bath has a volume of 2L, temperature up to 100°C, temperature uniformity 0.24°C at 37°C, stability 0.5°C at 37°C, 115 V. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Fisher Scientific Isotemp 102 Water Bath includes:
- > Lid
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific Isotemp 102 Water Bath Specifications
Temperature Range
|
Ambient to 100°C
|
Temperature Uniformity
|
±0.24°C at 37°C
|
Control
|
Analog
|
Capacity
|
2L (0.52 gal)
|
Bath Dimensions (L x W x H)
|
5.25 x 5.88 x 6 in (13.3 x 14.9 x 15.2 cm)
|
Electrical Requirements
|
120V, 60Hz, 1.7A
|
Temperature Stability
|
±0.5°C at 37°C
|
Overall Dimensions (L x W x H)
|
9.88 x 10.75 x 7.62 in (25.1 x 27.3 x 19.4 cm)
|
Click Here to View all the Technical Specs for the Isotemp 102
- Confirm power supply and PCB are outputting stable, correct voltages across all essential circuits to ensure reliable operation
- Inspect the display for any dim, flickering, or dead pixels, replacing any compromised sections to maintain full clarity and accurate readouts
- Verify that all buttons, knobs, and switches are functioning properly, providing precise and responsive control across all settings
- Conduct a thorough inspection of the water reservoir for any holes, rust, or other structural damage that may impact its integrity
- Fill the reservoir with water, checking closely for any external or internal leaks that could affect performance
- Examine the lid for any signs of wear or damage, replacing hinges and handles if necessary to ensure a secure and proper seal
- Inspect for any internal leaks originating from the water reservoir, confirming the unit’s internal plumbing integrity
- Conduct a comprehensive inspection of internal wiring for any breakages, tears, or wear; replace any underperforming or damaged components to maintain safe operation
- Examine internal insulation for signs of burns, breaks, or deterioration, replacing as needed to ensure thermal efficiency and safety
- Confirm that all feet are intact and that the bath sits level and balanced on a flat surface to prevent movement or instability
- Calibrate both temperature accuracy and uniformity using a NIST-traceable temperature probe, verifying compliance with specification standards
- Perform a detailed cosmetic inspection to confirm that all external surfaces and finishes meet aesthetic and quality standards
Technician Inspection and Recertification Checklist: Water Bath
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm power supply and PCB output stable, correct voltages across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace for clarity
|
Completed
|
|
Verify functionality of all buttons, knobs, and switches for responsive control
|
Completed
|
|
Inspect water reservoir for holes, rust, or structural damage
|
Completed
|
|
Fill reservoir with water; check for external or internal leaks
|
Completed
|
|
Examine lid for wear or damage; replace hinges or handles as needed
|
Completed
|
|
Inspect for internal leaks from water reservoir to confirm plumbing integrity
|
Completed
|
|
Check internal wiring for damage; replace underperforming components for safety
|
Completed
|
|
Examine internal insulation for burns, breaks, or deterioration; replace as needed
|
Completed
|
|
Confirm feet are intact and bath is balanced on flat surface
|
Completed
|
|
Calibrate temperature accuracy and uniformity with NIST-traceable probe
|
Completed
|
|
Perform cosmetic inspection to confirm quality of external surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|