Fisher Scientific 88861045 3D Platform Rotator
The Fisher Scientific 88861045 3D Platform Rotator offers precise and smooth orbital motion, making it an ideal choice for mixing in various laboratory vessels such as culture flasks, dishes, plates, and tubes. With a user-friendly interface, it allows for simple speed adjustment within the range of 2 to 80 RPM, ensuring accurate mixing with a speed accuracy of ±1 RPM. The rotator's gentle movement is well-suited for tasks like staining gels without causing damage to edges or maintaining the integrity of delicate tissues during incubation. Its large platform, equipped with a non-slip mat, can accommodate commonly used vessels, preventing sample movement.
Additionally, the low voltage power supply ensures safe operation in cold rooms while minimizing energy consumption. This versatile rotator can operate within a temperature range of 5°C to 40°C, making it suitable for use in refrigerators or non-CO2 incubators. It offers both continuous and timed operation options with automatic switch-off functionality. The LED display conveniently shows the last used settings, even after the power has been turned off. With a maximum load capacity of 1.8 lb. (0.8 kg) and a platform size measuring 12 x 10.5 inches (30.6 x 26.6 cm), the Fisher Scientific 88861045 3D Platform Rotator is a reliable and efficient solution for laboratory mixing needs.
The Fisher Scientific 88861045 3D Platform Rotator includes:
- > 12" x 10.5" Dimpled Rubber Mat Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific 88861045 3D Platform Rotator Specifications
Speed Range
|
2 to 80 RPM
|
Speed Accuracy
|
± 1 RPM
|
Tilt Angle
|
7°
|
Timer
|
1 minute to 99 hours 59 minutes
|
Display
|
LED
|
Maximum Load
|
1.8 lb. (0.8 kg)
|
Platform Size (L x W)
|
12 x 10.5 in. (30.6 x 26.6 cm)
|
Dimensions (L x W x H)
|
16.3 x 11.6 x 6.7 in. (41.5 x 29.5 x 17 cm)
|
Weight
|
16.1 lb. (7.3 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz, 0.8A
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|