Fisher Scientific 88861026 Open Air Rocker
The Fisher Scientific 88861026 Open Air Rocker is a versatile laboratory equipment ideal for various cell culture applications and environmental conditions. It offers precise control with a speed range of 1 to 50 RPM and an accuracy of ± 1 RPM. The adjustable rocking angle, ranging from 1° to 15° with a ± 1° accuracy, can be electronically controlled while the unit is in operation. This rocker features a user-friendly microprocessor control system with LED displays for speed, tilt angle, and time, ensuring repeatable and accurate results. Additionally, it comes equipped with safety features such as audible and visual signals to detect obstructions or overloads and a gradual speed increase to prevent splashing. This rocker can be used in refrigerators, cold rooms, or non-CO2 incubators, making it a reliable choice for various laboratory tasks, including staining, hybridization, hematology, and blotting techniques.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Fisher Scientific 88861026 Open Air Rocker includes:
- > 12.4" x 10.9" Dimpled Rubber Mat Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific 88861026 Open Air Rocker Specifications
Speed Range
|
1 to 50 RPM
|
Speed Accuracy
|
± 1 RPM
|
Angle Range
|
1° to 15°
|
Angle Accuracy
|
± 1°
|
Timer
|
1 second to 9999 minutes
|
Display
|
LED
|
Maximum Load
|
9.9 lb. (4.5 kg)
|
Rocking Motion
|
Left right reciprocating
|
Platform Size (L x W)
|
12.4 x 10.9 in. (31.6 x 27.6 cm)
|
Dimensions (L x W x H)
|
16.3 x 11.5 x 8 in. (41.5 x 29.3 x 20.4 cm)
|
Weight
|
30 lb. (12.7 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz, 0.1A, 13.5W
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|