The Fisher Scientific 88861023 Microplate Shaker is a versatile and essential tool for a wide range of applications in life sciences, immunoassays, and more. Designed to meet the needs of laboratory professionals, this microplate shaker offers precise and variable shaking speeds, ranging from 150 to 1000 RPM, with an impressive speed accuracy of ± 1% up to 299 RPM and ± 2% from 300 to 1000 RPM. One of its standout features is its user-friendly design, allowing for quick and easy fitting of up to 4 x 96-well microplates.
The shaker offers both continuous and timed operation, complete with an electronic timer that can be set for up to 99 hours and 59 minutes, ensuring accurate and repeatable time-sensitive runs. The shaker's display is equipped with LED technology, making it easy to monitor and control your experiments. It also boasts a memory function, displaying your last used settings even after power has been turned off, and seamlessly resuming operation at the same speed and remaining time when power is restored. This feature is especially valuable for uninterrupted and consistent experiments.
The Fisher Scientific 88861023 Microplate Shaker includes:
- > 10.8" x 9.3" Microplate Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific 88861023 Microplate Shaker Specifications
Speed Range
|
150 to 1,000 RPM
|
Speed Accuracy
|
± 1% of set speed up to 299 RPM
± 2% from 300 to 1000 RPM
|
Timer
|
0 minutes to 99 hours 59 minutes
|
Orbit
|
2.5 mm (0.098 in.)
|
Display
|
LED
|
Maximum Load
|
2.2 lb. (1 kg)
|
Microplate Capacity
|
4 x 96-well microplates
|
Platform Size (L x W)
|
10.8 x 9.3 in. (27.5 x 23.5 cm)
|
Dimensions (L x W x H)
|
16.3 x 11.5 x 5.4 in. (41.5 x 29.3 x 14.3 cm)
|
Weight
|
25.3 lb. (11.5 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|