Eppendorf Thermostat Plus Microplate Heater, Catalog # 5352
The Eppendorf ThermoStat Plus Interchangable Block Heater is
perfect for cell technology, enzymatic testing, molecular biology,
immunological testing and a large array of general laboratory applications. The
onboard digital display allows the user to easily select both time and
temperature with little to no effort. Programs can be entered, stored, and
started with a simple press of a button. With an extreme temperature accuracy of
±1 at 0, ±0.5 at 37, ±1 at 90°C and an above average temperature range of 30
below room temperature to 99°C, the Thermostat Plus can provide both reliability
and versatility for almost any laboratory situation. An extreme heat up rate of
4 degrees Celsius per minute allows for speedy testing periods and high overall
throughputs. This block heater also provides users with the ability to save up
to 4 different temperature settings for maximum user efficiency.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Eppendorf Thermostat Plus Microplate Heater, Catalog # 5352 includes:
- > MTP Block & Lid
- > Operator's Manual
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf Thermostat Plus Microplate Heater 5352 Specifications
Temperature Range
|
30 below room temperature to 99°C
|
Temperature Accuracy
|
±1 at 0, ±0.5 at 37, ±1 at 90°C
|
Height x Width
|
4.9 inches x 8.7 inches
|
Display
|
Digital
|
Timer
|
1 min to 99:59
|
Weight
|
11 pounds
|
Eppendorf Thermostat Plus ManualClick Here to View all the Technical Specs for the Thermostat Plus
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|