Eppendorf 6321 Mastercycler Pro Gradient Thermal Cycler
The Eppendorf 6321 Mastercycler Pro Gradient Thermal Cycler is a state-of-the-art device designed for efficient and precise PCR protocols. Offering a thermal range of 4°C to 99°C for its block and 37°C to 110°C for the lid, the system ensures high accuracy and uniformity in temperature control. It boasts a rapid heating rate of 4°C/sec and cooling rate of 3°C/sec thanks to the Triple Circuit Technology, ensuring quick yet even temperature distribution. Its SteadySlope Technology ensures that the heating and cooling ramp rates for thermal modules are identical in both gradient and normal operation. Users can expect identical temperature control characteristics in both optimization and routine experiments, as well as a reliable transfer of the optimization results to the routine application.
The Mastercycler Pro has a sample capacity of 96 x 0.2 ml PCR tubes or a 96-well PCR plate and provides memory storage for up to 700 programs organized into 100 folders. One of its standout features is the vapo.protect heated lid, which evenly applies pressure to tubes or plates, effectively minimizing evaporation. This coupled with its TSP technology ensures optimal sample protection by maintaining the thermoblock at 20°C during the lid heating phase, thus reducing non-specific product formation risks. Users can also utilize the built-in self-test function to ensure the thermoblock's proper functionality, and it’s conveniently controlled via a dedicated control panel or the PC software CycleManager pro. With its compact dimensions, robust design, and advanced features, this thermal cycler is ideal for both routine and specialized PCR applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf 6321 Mastercycler Pro Gradient Thermal Cycler includes:
- > 96 Well Block
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf 6321 Mastercycler Pro Gradient Thermal Cycler Specifications
Thermal Range (Block)
|
4°C to 99°C
|
Thermal Range (Lid)
|
37°C to 110°C
|
Thermal Accuracy
|
± 0.2°C
|
Thermal Uniformity
|
≤ ± 0.3°C at 35°C
≤ ± 0.4°C at 90°C
|
Ramping Rates
|
Heating: 4°C/sec
Cooling: 3°C/sec
|
Sample Capacity
|
96 x 0.2 ml PCR tubes or 96-well PCR plate
|
Memory
|
Up to 700 programs and 100 folders
|
Maximum Number of Cycles
|
99 (1 to 20 temperature control steps per cycle)
|
Gradient Range
|
1°C to 20°C
|
Gradient Temperature Range
|
30°C to 99°C
|
Connections
|
2 x USB and 1 x MMC slot
|
Dimensions (W x D x H)
|
10.2 x 21.9 x 14.6 in. (26 x 55.5 x 37 cm)
|
Weight
|
44.5 lb. (20.2 kg.)
|
Electrical Requirements
|
100-130V, 50/60Hz, 950W
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|