Eppendorf 5920R Refrigerated Centrifuge
Temperature Range
-11°C to 40°C
- Swing-bucket rotors and adapters accommodate tubes and bottles from 0.2 mL to 1,000 mL
- Plate rotor options for centrifugation of all types of MTP, PCR or Deepwell Plates
- Fixed-angle rotors for high-speed molecular biology applications in tubes from 0.2 mL to 50 mL
- High centrifugation speed of up to 21,194 × g (13,700 rpm)
- Automatic rotor recognition and imbalance detection for maximum operational safety
- Temperature range from -11 °C to 40 °C
- FastTemp function for quick pre-cooling
- FastTemp pro® function allows for automatic precooling based on pre-programmable time and date
- Continuous cooling maintains a constant temperature after the run has been completed – your samples stay cool
- ECO shut-off engages after 8 hours (adjustable) of non-use to reduce energy consumption and to extend compressor life
- Dynamic compressor control (DCC) technology for precise cooling performance
The Eppendorf 5920R Refrigerated Centrifuge is a large benchtop refrigerated centrifuge with exceptional performance and high capacity, making it a superior choice for high-throughput applications. The Eppendorf 5920R combines outstanding functionality with an advanced operating system. It has a menu-driven, multilingual operation menu with a large backlit display that allows the user to save up to 99 programs. The FastTemp Pro function is also featured, providing fast, automated pre-cooling using a pre-programmable time and date. The refrigeration system also utilizes Dynamic Compressor Control (DCC) technology for precise cooling performance.
The Eppendorf 5920R Refrigerated Centrifuge has a wide variety of rotors, buckets, and adapters available to spin all commonly used tubes, bottles, and plates. With high centrifugation speeds of up to 13,700 RPM (21,194 x g), the 5920R is certainly capable of suiting a large multitude of different applications. This centrifuge also features automatic rotor recognition along with imbalance detection for maximum operational safety. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
(This centrifuge is in like-new condition out of box)
The Eppendorf 5920R Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5920R Refrigerated Centrifuge Specifications
Maximum Speed
|
13,700 RPM
|
Maximum RCF
|
21,194 x g
|
Maximum Capacity
|
4 x 1000ml
|
Temperature Range
|
-11°C to 40°C
|
Timer
|
10 seconds to 99:59 hours, continuous
|
Acceleration Rate Settings
|
9
|
Braking Rate Settings
|
9 or no brake
|
Maximum Noise Level
|
< 67 dB(A)
|
Dimensions (W x D x H)
|
29.2 x 27.8 x 15.9 in (74 x 71 x 40 cm)
|
Weight
|
306 lb.
|
Electrical Requirements
|
120V, 50/60Hz, 12A
|
Click Here to Expand Rotor Specifications
S-4xUniversal Rotor Specs
|
Capacity
|
4 x 750ml
|
Max RPM
|
4,250
|
Max RCF
|
4,402 x g
|
S-4-1000 Rotor Specs
|
Capacity
|
4 × 1000ml
|
Max RPM
|
3,700
|
Max RCF
|
3,428 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|