Eppendorf 5910 Ri Refrigerated Centrifuge
Maximum Speed 14,000 RPM (Fixed Angle Rotor)
Temperature Range
-11°C to 40°C
Maximum Capacity
4 x 1000ml
- Large 7-inch VisioNize touch interface for intuitive operation and real-time monitoring
- Favorites function for quick access to frequently used settings directly from the home screen
- Up to 99 programmable runs with customizable parameters, rotor compatibility check, and ramp settings
- Advanced Dynamic Compressor Control (DCC) technology for precise and efficient cooling
- Automatic rotor recognition and lifetime tracking for enhanced safety and reliability
- Built-in imbalance sensors and maintenance reminders to minimize downtime
- Quiet operation with noise levels below 59 dB(A) for a user-friendly lab environment
- Compact and ergonomic design with dimensions of 28.3 x 25.9 x 14.5 in for efficient use of space
- Documentation function with user management, password protection, and activity logs for enhanced quality control
The Eppendorf 5910 Ri Refrigerated Centrifuge combines high capacity and performance in a compact footprint with an ergonomic design, standing out for its ease of use and versatility. This centrifuge features a large 7-inch VisioNize touch interface with an intuitive user experience providing fast and error-free operation, ensuring efficient and reproducible runs. Additionally, the documentation function and unique user management support your quality control with multiple authorization levels, password protection, and documentation of all runs and user activities. With the unique favorites function, the setting of parameters is now faster than ever. Frequently-used settings for time, temperature, and speed are directly accessible on the home screen, allowing users to create individual runs with just three clicks. With the program function, up to 99 programs each consisting of all run parameters (incl. ramps and radius correction settings) with clearly assigned names and a rotor compatibility check can be saved to avoid operation errors. The advanced temperature management of the Eppendorf 5910 Ri Refrigerated Centrifuge is more than just cooling. It offers the latest technology, such as quick pre-cooling and our patented Dynamic Compressor Control (DCC) technology for precise cooling performance. This centrifuge features an overall temperature range of -11°C to 40°C, along with a maximum speed of up to 14,000 RPM with the fastest available fixed angle rotor. Contemporary safety features such as automatic rotor recognition, rotor lifetime tracking, the new rotor compatibility check, built-in imbalance sensors, and much more ensure peace of mind. Freely programmable reoccurring tasks, such as maintenance reminders or rotor and device cleaning schedules, help to optimize maintenance of your centrifuge, reducing downtime and therefore enhancing the efficiency of the user’s lab. Together with a comprehensive portfolio of fixed-angle and swing-bucket rotors, this centrifuge accommodates a particularly broad spectrum of applications, making it the perfect choice for multiuser labs with diverse requirements. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Eppendorf 5910 Ri Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5910 Ri Refrigerated Centrifuge Specifications
Maximum Speed
|
14,000 RPM
|
Maximum RCF
|
22,132 x g
|
Maximum Capacity
|
4 x 1000ml
|
Temperature Range
|
-11°C to 40°C
|
Timer
|
10 seconds to 99 hours 59 minutes, continuous run (hold), or short-spin
|
Acceleration Rates
|
10
|
Deceleration Rates
|
10
|
Noise Level
|
< 59 dB(A)
|
Display
|
VisioNize touch interface
|
Dimensions (W x D x H)
|
28.3 x 25.9 x 14.5 in. (72 x 68 x 37 cm)
|
Weight
|
240.3 lb. (109 kg.)
|
Electrical Requirements
|
120V, 60Hz, 12A
|
Click Here to Expand Rotor Specifications
S-4xUniversal Rotor Specs
|
Capacity
|
4 x 750ml
|
Max RPM
|
4,250
|
Max RCF
|
4,402 x g
|
S-4-1000 Rotor Specs
|
Capacity
|
4 × 1000ml
|
Max RPM
|
3,700
|
Max RCF
|
3,428 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|