Eppendorf 5810R Refrigerated Centrifuge (20 AMP)
Maximum Speed 14,000 RPM (Fixed Angle Rotor)
Temperature Range –9 °C to 40°C
Maximum Capacity 4 x 750ml
- Brushless motor with high reliability.
- Internal capacity for up to 750 ml in tubes or 16 microplates.
- Accommodates various rotors including F-45-30-11, A-4-81, and A-4-62.
- Maximum speed of 14,000 RPM and a maximum g-force of 20,800 x g.
- Compact design with a height of only 13 inches, suitable for limited spaces.
- Dynamic temperature range from -9°C to 40°C.
- Onboard LCD display for rotation speed, time left in the cycle, and internal temperature.
- Increased power of 20A model allows for lower temperatures at maximum centrifugation speed and quicker cooling down to the set temperature.
The Eppendorf 5810R Refrigerated Centrifuge (20 AMP) features a brushless motor and an internal capacity of up to four 400 ml tubes or 16 microplates. This centrifuge accommodates a variety of rotors to meet your application needs including the F-45-30-11 Rotor, the A-4-81 Rotor, and the A-4-62 Swing bucket rotor and more. This versatile centrifuge has a maximum speed of up 14,000 rotations per minute and a maximum g-force of 20,800 x g, making it the perfect benchtop centrifuge for just about any application. With a compact height of only 13 inches, the 5810R is perfect for use in close areas or under storage shelving. This Eppendorf centrifuge has dynamic temperate capability ranging from as cold as -9°C to as hot as 40°C. With the increased power of this 20A model, lower temperatures are attainable at maximum centrifugation speed in comparison to the 15A model. This increased power also allows for quicker cooling down to the set temperature. The 5810R is considered a stable centrifuge throughout the scientific community and can be used in applications ranging from separating cell lysates to creating DNA precipitations. The refrigerated centrifuge also features an onboard LCD display for rotation speed, time left in cycle, and internal temperature.
We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire. Please Note: This is the 20A model of the 5810R which has a special 120V/20A mains connection with the power cord permanently fitted to the centrifuge.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Eppendorf 5810R Refrigerated Centrifuge (20 AMP) includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Fixed Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5810R Refrigerated Centrifuge (20 AMP) Specifications
Maximum Speed
|
14,000 RPM (Fixed Angle Rotor)
|
Maximum g-Force
|
20,800 x g(RCF)
|
Temperature Range
|
–9 °C to 40°C
|
Power
|
120 V/60 Hz
|
Depth
|
24” in
|
Width
|
28” in
|
Height
|
13” in
|
Weight
|
218 lbs
|
Click Here to Expand Rotor Specifications
A-4-62 Rotor Specs
|
Capacity
|
4 round buckets up to 250ml per
|
Max RPM
|
4,000
|
Max RCF
|
3,220 x g
|
A-2-DWP-AT Rotor Specs
|
Capacity
|
2 × 6 MTP
|
Max RPM
|
4,500
|
Max RCF
|
3,486 x g
|
A-4-81 Rotor Specs
|
Capacity
|
4 × 500 ml
|
Max RPM
|
4,000
|
Max RCF
|
3,220 x g
|
F-34-6-38 Rotor Specs
|
Capacity
|
6 × 85 mL tubes
|
Max RPM
|
12,000
|
Max RCF
|
18,514 x g
|
F-45-30-11 Rotor Specs
|
Capacity
|
30 × 1.5/2.0 mL tubes
|
Max RPM
|
14,000
|
Max RCF
|
20,817 x g
|
FA-45-48-11 Rotor Specs
|
Capacity
|
48 × 5 mL tubes
|
Max RPM
|
13,100
|
Max RCF
|
20,913 x g
|
S-4-104 Rotor Specs
|
Capacity
|
4 × 750 mL
|
Max RPM
|
3,900
|
Max RCF
|
3,214 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|