Eppendorf 5427R Refrigerated Centrifuge
Temperature Range -11°C to 40°C
Maximum Capacity 48 x 1.5/2.0ml
- Designed for high sample throughput, addressing capacity limitations found in other models
- Offers rapid run times combined with reliable performance for efficient sample processing
- User-friendly design with an intuitive interface for straightforward adjustments
- Features innovative compressor technology for eco-friendly, precise temperature control
- Compatible with a wide array of high-quality aluminum rotors tailored for specific applications
- Timer settings range from 10 seconds to 9:59 hours, including a continuous run option
- Low noise operation with a level under 56 dB(A) ensures a quieter lab environment
- User-friendly design with an intuitive interface for easy operation
The Eppendorf 5427R Refrigerated Centrifuge 5427R has an aerosol-tight 48-place rotor
increases productivity and can handle though challenges in a safe and reliable
manner at a maximum speed of 25,000 x g. The rotors are equipped with the Eppendorf
Quick Lock lids for ease of opening and closing the rotor lid requiring only a
quart turn. Plus, for operational safety there is automatic rotor recognition
and imbalance detection. The lid will lock automatically with a soft-touch,
one-finger ergonomic approach. This centrifuge has temperature range of -11 °C
to 40 °C and a FastTemp feature that allows for quick pre-cooling. A
condensation drain is built-in to eliminate water accumulation. You can program
the ECO shut-off to engage after 1, 2, 4, or 8 hours of non-use.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Eppendorf 5427R Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > Power Cord
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Hex Rotor Wrench
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5427R Centrifuge Specifications
Speed Range
|
100-16,220 RPM
25,000 RCF
|
Max. capacity
|
48 × 1.5/2.0 mL
|
Acceleration time
|
18 s
|
Deceleration time
|
18 s
|
Display
|
Large LCD
|
Timer
|
10 s to 9:59 h
|
Noise level
|
< 46 dB(A) with Rotor F-45-48-11
|
Volume range
|
96 mL
|
Max Speed
|
16,220 RPM
|
Power Supply
|
50-60 Hz
|
Max. Power Consumption
|
550 W
|
Dimensions (W × D × H)
|
31.9 × 54.0 × 25.4 cm
|
Weight w/o accessories
|
30 kg
|
Cooling
|
Refrigerated
|
Temperature control range
|
-11 °C to 40 °C
|
Click Here to Expand Rotor Specifications
FA-45-24-11 Rotor Specs
|
Capacity
|
24 x 1.5/2.0 ml tubes
|
Max RPM
|
16,220
|
Max RCF
|
25,000 x g
|
FA-45-30-11 Rotor Specs
|
Capacity
|
30 x 1.5/2.0 ml tubes
|
Max RPM
|
14,000
|
Max RCF
|
20,800 x g
|
FA-45-48-11 Rotor Specs
|
Capacity
|
48 x 1.5/2.0 ml tubes
|
Max RPM
|
12,700
|
Max RCF
|
18,210 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|