Eppendorf 5418R Refrigerated Centrifuge w/ FA-45-18-11 Rotor & Lid
Speed Range 100 to 14,000 RPM
Max. Capacity 18 x 1.5/2.0ml
Temperature Range 0°C to 40°C
- Designed for molecular biology and clinical applications, accommodates 18 x 1.5/2.0 ml tubes
- Reaches speeds up to 14,000 RPM with a maximum RCF of 16,873 x g for efficient processing
- Compact design with a small footprint and low access height, ideal for limited spaces
- Operates quietly at noise levels below 57 dB(A), ensuring a comfortable lab environment
- Features soft-touch one-finger lid closure and Eppendorf QuickLock for secure, easy locking
- Temperature control from 0°C to 40°C with a FastTemp function for quick pre-cooling
- ECO power shut-off enhances energy efficiency by powering down when not in use
- Adjustable run times from 0.5 to 99 minutes, or indefinitely, for flexible protocol management
The Eppendorf 5418R Refrigerated Centrifuge epitomizes efficiency and precision, tailored for the demanding environments of molecular biology and clinical applications. It boasts a robust capacity of 18 x 1.5/2.0 ml tubes, reaching maximum speeds of 14,000 RPM and achieving a remarkable maximum RCF of 16,873 x g. Its design is uniquely compact, offering a minimal footprint and low access height that make it ideal for laboratories constrained by space. Despite its power, the 5418R operates under a noise level of less than 57 dB(A), ensuring a non-disruptive, comfortable lab environment.
Further enhancing its user-friendly features, this centrifuge includes a soft-touch, one-finger lid closure and the Eppendorf QuickLock lid, which allows secure, effortless locking with just a quarter turn. Temperature control ranges from 0°C to 40°C, facilitated by a FastTemp function that swiftly cools the unit to the desired temperature. The centrifuge also supports environmental conservation with its ECO power shut-off feature, which reduces energy consumption by powering down the unit when not actively in use. Moreover, it offers flexible run times from 0.5 minutes to 99 minutes or indefinitely, accommodating a variety of protocols and user preferences.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf 5418R Refrigerated Centrifuge includes:
- > FA-45-18-11 Rotor & Lid
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Rotor Wrench
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5418R Refrigerated Centrifuge w/ FA-45-18-11 Rotor Specifications
Speed Range
|
100 to 14,000
|
Max RCF
|
16,873 x g
|
Max Capacity
|
18 x 1.5/2.0ml
|
Temperature Range
|
0°C to 40°C
|
Acceleration Time
|
13 s
|
Deceleration Time
|
13 s
|
Run Time
|
0.5 min to 99 min, as well as infinity (oo),
Adjustable to 10 min in 0.5 min increments,
then increments of 1 min
|
Dimensions (W x D x H)
|
298 x 463 x 463 cm (11.73 x 18.23 x 18.23 in)
|
Electrical Requirements
|
120V, 50/60Hz, 320W
|
Noise Level
|
< 57 dB (A)
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|