Eppendorf 5415D Digital Centrifuge
Speed Range Variable Speed up to 13,200 RPM
Maximum Capacity 24 x 1.5ml Tubes
- Easy-to-use with knob control, ideal for routine tasks.
- Turn-key, walk-away operation for convenience.
- Maintenance-free with a brushless motor.
- Maximum speed of 13,200 rpm for efficient centrifugation.
- Digital timer with a run time of up to 30 minutes.
- Capacity for up to 24 x 1.5ml tubes (or 36 x 0.5ml).
- Space-saving design suitable for tabletop use.
- Fully tested for performance at factory specifications before shipping.
The Eppendorf 5415D
digital table top microfuge is extremely easy to use with knob control, virtually working as a turn key, walk away machine. Its small footprint and
brushless motor conveniently make it maintenance free. The 5415D
has a maximum speed of 13,200 rpm and has a digital timer run time of up to 30
minutes. The Eppendorf 5415D has a maximum capacity of up to 24 x 1.5ml tubes (or 36 x 0.5ml). With its space-saving design and user-friendly operating concept, the centrifuge is ideal for everyday routine tasks.
All of our equipment is fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Eppendorf 5415D Digital Centrifuge includes:
- > 24-Place 1.5ml F-45-24-11 Rotor with Lid
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5415D Digital Centrifuge Specifications
Maximum Speed
|
Variable speed up to 13,200 rpm
|
Maximum g –Force (RCF)
|
16,110 x g
|
Maximum capacity
|
Rotor: 24 x 1.5ml tubes
|
Acceleration Time
|
< 15 sec
|
Deceleration Time
|
< 13 sec
|
Digital Timer
|
Up to 30 minutes
|
Dimensions (H x D x W)
|
230 x 310 x 230 mm
|
Power
|
110-115 volt 60Hz
|
Refrigeration
|
No Refrigeration
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
Eppendorf 5415D Centrifuge with F-45-24-11 Rotor
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|