Eppendorf 5390 Microplate Heat Sealer
The Eppendorf 5390 Microplate Heat Sealer is an ideal solution for effectively sealing plates crafted from diverse plastics, designed to cater to applications with low to medium throughput requirements. Its feature of pre-configured temperature controls ensures the achievement of the best possible sealing outcomes. When used together with Eppendorf's specialized heat sealing films and foils, it creates an almost perfect seal, safeguarding against sample loss through exposure to elevated temperatures or during extended storage periods. This system is especially beneficial for maintaining sample integrity and minimizing potential compromises in various storage conditions. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf 5390 Microplate Heat Sealer includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5390 Microplate Heat Sealer Specifications
Internal Heater Temperature Range
|
Ambient to 220°C
|
Heater Plate Maximum Temperature
|
~ 170°C
|
Material
|
Sheet Metal
|
Dimensions (W x D x H)
|
5.5 x 5.9 x 12.9 in. (14 x 15 x 33 cm)
|
Weight
|
7.5 lb. (3.4 kg.)
|
Electrical Requirements
|
115V, 47-63Hz, 500W
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|