Eppendorf Vacufuge Plus Concentrator Complete System
- 4 Heating Levels Ambient, 30°C, 45°C, 60°C
- Max. Speed of 1,400 RPM
- Built-in Diaphragm Pump
- Full 180 Day Parts and Labor Warranty
Go to Product Details
Eppendorf 5301 Vacufuge Concentrator System
Heating Levels Ambient, 30°C, 45°C, 60°C
- Specifically designed for gentle concentration of DNA, RNA, nucleotides, proteins, and gel electrophoresis samples
- Combines centrifugal force with vacuum to maximize efficiency in sample processing
- Features three aeration modes: aqueous, alcohol, and off, for optimized evaporation
- Capable of reducing processing time by up to 20%, enhancing laboratory workflow
- Reaches a maximum speed of 1,400 RPM and a maximum RCF of 240 x g
- Offers four heating levels: Ambient, 30°C, 45°C, and 60°C, for versatile sample handling
- Includes a broad time range setting from 1 minute to 9 hours and 59 minutes, plus a continuous mode
The Eppendorf 5301 Vacufuge Concentrator System is in excellent condition. The Vacufuge combines centrifugal force and vacuum for gentle concentration of DNA, RNA, nucleotides, proteins, and samples for gel electrophoresis. Three aeration modes, including aqueous, alcohol, and off, maximize evaporation efficiency, prolong pump lifespan, and reduce processing time by up to 20%. Reaches maximum speeds up to 1,400 RPM and maximum RCF of 240 x g at a choice of four heating levels, including ambient, 30°C, 45°C, and 60°C. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf 5301 Vacufuge Concentrator System includes:
- > F-45-48-11 Fixed Angle Rotor
- > Built in Pump
- > Operator's Manual
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5301 Vacufuge Concentrator System Specifications
Max. Speed
|
1,400 RPM
|
Heating Levels
|
Ambient, 30°C, 45°C, 60°C
|
Time Range
|
1 min to 9 hr 59 min, continuous
|
Vacuum Level
|
< 20 mbar
|
Electrical Requirements
|
115V, 60Hz, 370W
|
Rotor Capacity
|
48 x 1.5/2.0 ml Tubes
|
Operator's Guide
Click Here to View all the Technical Specs for the Vacufuge
Eppendorf Vacufuge Plus Concentrator Complete System
- 4 Heating Levels Ambient, 30°C, 45°C, 60°C
- Max. Speed of 1,400 RPM
- Built-in Diaphragm Pump
- Full 180 Day Parts and Labor Warranty
Go to Product Details
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|