Corning PC 410 Stirrer
The Corning PC 410 Stirrer features dial controls for stir, an indicator light for power and a 5" X 7" platform. Stirring speeds from 60 to 1,100 RPM. Power: 120V All of our equipment will be fully tested to perform at factory specifications before being shipped.
Included with the Corning PC 410 Stirrer:
> Power Cord > Operator's Guide > Full 180 Day Parts and Labor Warranty
Corning PC 410 Stirrer Specifications
Stirrer Speed
|
60 to 1,100 RPM
|
Platform Size
|
5 x 7 in (12.7 x 17.8 cm)
|
Electrical Requirements
|
120V, 60Hz, 0.7A, 73W
|
Operator's Manual
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for clarity and replace if needed | Completed | |
Examine external regulators for integrity and damage | Completed | |
Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
Check internal wiring for frays, tears, or breakages | Completed | |
Confirm software settings match internal component specs | Completed | |
Inspect and clean fan, replacing if wear is detected | Completed | |
Verify condition of wheels and leveling mounts | Completed | |
Inspect door gaskets and latches for sealing integrity | Completed | |
Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
Confirm 24-hour temperature and CO₂ stability | Completed | |
Conduct cosmetic inspection to ensure appearance standards | Completed | |