|
|
|
|
Buchi R-200 Rotavapor System (No Pump)
The Buchi Rotavapor R-200 system has the ability to process
samples ranging in size from 50 mL to 3 liters and has a relatively small
footprint for just about any laboratory workspace. With a two piece cold trap condenser
the Buchi Rotavapor R-200 system is perfect for low boiling solvents. This rotavapor system also features an internal motorized hydraulic mechanism to
assist users with the raising and lowering of samples and compounds. A digital display
exhibits all necessary data to the user at all times with sharp visibility. This rotavapor system does not include a vacuum pump.
Includes: a V-800 for optimal vacuum precision and complete
control over compound elements such as temperature and pressure levels. Along
with the V-800, this R-200 Rotavapor system Includes: a Buchi B-490 Water bath
with temperature range of +20°C to +180°C, a digital display and a large liquid
capacity to accommodate just about any sample flask. To help facilitate compound evaporation the R-200
Rotavap also Includes: A two-piece cold trap condenser, a
motorized hydraulic mechanism, a visible LCD display, and a slew of devices to
help facilitate compound evaporation the Buchi Rotavapor R-200 System is
perfect for almost all applications.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Buchi R-200 Rotavapor System (No Pump) includes:
- > B-490 water bath
- > V-800 vacuum controller
- > Rotavap “C” glassware attachments
- > Wouff Bottle and Valve Controller
- > 1 Liter Evaporating Flask
- > 500ml Receiving Flask
- > All essential Vacuum Tubing & Power Cords
- > Full 180 Day Parts and Labor Warranty
Buchi R-200 Rotavapor System (No Pump) Specifications
Rotavapor R-200
|
AC input
|
120 V, 50 – 60 Hz
|
Speed
|
20 -280 rpm
|
Weight
|
16 -18 kilograms
|
Temperature Range
|
+20°C to +180°C
|
Condenser
|
Two-piece cold trap condenser
|
Capacity Range
|
50 mL – 3L
|
Counter Space Footprint
|
0.14 m²
|
V-800 Vacuum Controller
|
Vacuum Reading Range
|
0 to 1400 Millibar
|
Hysteresis
|
Automatic or 1 – 500 mbar
|
Regulating range
|
1 mbar/torr to atmospheric pressure
|
Display
|
Digital LCD Display
|
Power Consumption
|
16 watts
|
B-490 Heating Bath
|
Display
|
LCD Digital Display
|
Control
|
Analog Control Knobs
|
Temperature range
|
20 to 180°C
|
Temperature Accuracy
|
+ 1ºC
|
- Inspect all glassware for chips or cracks, reapplying vacuum grease to stopcocks to ensure a secure seal
- Examine displays, replacing any dim or dead pixels to maintain clear visibility
- Verify drive motor RPM using a tachometer for accurate rotational speed
- Confirm the ease of raising and lowering the tower, lubricating as needed for smooth operation
- Check the Woulff bottle for any leaks or structural breakages, ensuring airtight performance
- Disassemble and inspect the vacuum pump, cleaning and replacing seals or components as necessary
- Inspect internal wiring for each component, checking for any frays, cuts, or potential breakages
- Examine all hoses, hose barbs, and seals for signs of wear or leaks, ensuring integrity
- Inspect the vapor duct seal and replace if compromised to maintain vacuum efficiency
- Verify and calibrate the vacuum controller for accurate pressure regulation
- Confirm all hoses and cables are securely fastened to prevent disconnections or leaks
- Check heat bath temperatures against settings to ensure precise thermal control
- Confirm the final system vacuum level meets manufacturer specifications
- Ensure cosmetic condition meets standards, with no damage impacting function or appearance
Technician Inspection and Recertification Checklist: Rotary Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect all glassware for chips or cracks, apply vacuum grease to stopcocks
|
Completed
|
|
Examine displays and replace dim/dead pixels
|
Completed
|
|
Verify drive motor RPM with tachometer
|
Completed
|
|
Confirm ease of raising/lowering tower, lubricate if needed
|
Completed
|
|
Check Woulff bottle for leaks or structural issues
|
Completed
|
|
Disassemble and inspect vacuum pump, clean/replace parts as needed
|
Completed
|
|
Inspect internal wiring for frays, cuts, or breakages
|
Completed
|
|
Examine hoses, hose barbs, and seals for leaks
|
Completed
|
|
Inspect vapor duct seal, replace if compromised
|
Completed
|
|
Verify and calibrate vacuum controller
|
Completed
|
|
Confirm hoses and cables are securely fastened
|
Completed
|
|
Check heat bath temperatures for accurate control
|
Completed
|
|
Confirm final system vacuum level meets specifications
|
Completed
|
|
Ensure cosmetic condition meets standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|