Buchi R-100 Rotavapor System with Chiller
The Buchi R-100 Rotavapor System with Chiller is an advanced solution for high-speed, single-stage distillations with minimal stress on samples. It includes the versatile Buchi B-100 Heating Bath, which offers digital temperature control from 20°C to 95°C and a 4L capacity. The system’s modular design allows for customization with various glassware assemblies, enhancing its adaptability for different laboratory needs. Key features of the R-100 Rotavapor include a rotation speed range of 20 to 280 rpm and a pressure tolerance of up to 6 bar on the condenser coil. The accompanying V-100 Vacuum Pump delivers impressive performance with a final vacuum of less than 10 mbar and a pumping volume of 1.5 m³/h, along with a power consumption of 150W. This pump, combined with the system's multifunctional valve unit, ensures precise vacuum control and efficient operation.
The system is also equipped with the Buchi I-100 Interface, enhancing vacuum control and monitoring with a measurement range of 0 to 1,400 mbar and accuracy of ± 2 mbar post-calibration. It features automatic hysteresis adjustment or manual setting from 1 to 500 mbar, with a regulating range of 0 to 1,100 mbar, and consumes 30W of power. To complement the Rotavapor system, the Brand-New Polyscience LM61GX1A110C LM6 Chiller, 560W provides reliable cooling with a temperature range of -10°C to 40°C and accuracy of ± 0.5K. It offers a cooling capacity of 350W and a maximum flow rate of 9L/min (2.67 gpm).
All of our equipment is fully tested to perform at factory specifications before being shipped.
TheBuchi R-100 Rotavapor System with Chiller includes:
- > Coil Glassware (Dry Ice Condenser is Pictured)
- > 1 Liter Evaporating Flask
- > 1 Liter Receiving Flask
- > Brand-New Polyscience LM61GX1A110C LM6 Chiller, 560W
- > Buchi B-100 Heating Bath
- > V-100 Diaphragm Pump
- > I-100 Interface Module
- > Vacuum Tubing and Connectors
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Buchi R-100 Rotavapor System with Chiller Specifications
Buchi R-100 Rotavapor System Specifications
|
Rotation Speed
|
20 to 280 rpm
|
Temperature
Control
|
20 ° to 95°C
|
Bath Capacity
|
4L
|
Pressure Load
on Condenser Coil
|
Max 6 bar
|
V-100 Vacuum Pump Specifications
|
Final Vacuum Possible
|
< 10 mbar
|
Pumping Volume
|
1.5mᶟ/h
|
Power Consumption
|
150 W
|
B-100 Heating Bath Specifications
|
Control
|
Analog Control Knobs
|
Temperature Range
|
20 to 95
°C
|
Temperature Accuracy
|
+ 1°C
|
Buchi I-100 Interface Specifications
|
Vacuum Measurement Range
|
0 to 1,400 mbar
|
Vacuum Measurement Accuracy
|
± 2 mbar (after calibration at constant temperature)
|
Hysteresis
|
Automatic or 1 to 500 mbar
|
Regulating Range
|
0 to 1,100 mbar
|
Power Consumption
|
30W
|
Polyscience LM61GX1A110C LM6 Chiller Specifications
|
Temperature Range
|
-10°C to 30°C
|
Temperature Stability
|
± 0.1°C
|
Cooling Capacity
|
560W
|
Maximum Flow Rate
|
7.9L/min (2.1 gpm)
|
Electrical Requirements
|
120V, 60Hz, 12A, 1Ph
|
- Inspect all glassware for chips or cracks, reapplying vacuum grease to stopcocks to ensure a secure seal
- Examine displays, replacing any dim or dead pixels to maintain clear visibility
- Verify drive motor RPM using a tachometer for accurate rotational speed
- Confirm the ease of raising and lowering the tower, lubricating as needed for smooth operation
- Check the Woulff bottle for any leaks or structural breakages, ensuring airtight performance
- Disassemble and inspect the vacuum pump, cleaning and replacing seals or components as necessary
- Inspect internal wiring for each component, checking for any frays, cuts, or potential breakages
- Examine all hoses, hose barbs, and seals for signs of wear or leaks, ensuring integrity
- Inspect the vapor duct seal and replace if compromised to maintain vacuum efficiency
- Verify and calibrate the vacuum controller for accurate pressure regulation
- Confirm all hoses and cables are securely fastened to prevent disconnections or leaks
- Check heat bath temperatures against settings to ensure precise thermal control
- Confirm the final system vacuum level meets manufacturer specifications
- Ensure cosmetic condition meets standards, with no damage impacting function or appearance
Technician Inspection and Recertification Checklist: Rotary Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect all glassware for chips or cracks, apply vacuum grease to stopcocks
|
Completed
|
|
Examine displays and replace dim/dead pixels
|
Completed
|
|
Verify drive motor RPM with tachometer
|
Completed
|
|
Confirm ease of raising/lowering tower, lubricate if needed
|
Completed
|
|
Check Woulff bottle for leaks or structural issues
|
Completed
|
|
Disassemble and inspect vacuum pump, clean/replace parts as needed
|
Completed
|
|
Inspect internal wiring for frays, cuts, or breakages
|
Completed
|
|
Examine hoses, hose barbs, and seals for leaks
|
Completed
|
|
Inspect vapor duct seal, replace if compromised
|
Completed
|
|
Verify and calibrate vacuum controller
|
Completed
|
|
Confirm hoses and cables are securely fastened
|
Completed
|
|
Check heat bath temperatures for accurate control
|
Completed
|
|
Confirm final system vacuum level meets specifications
|
Completed
|
|
Ensure cosmetic condition meets standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|