Buchi R-210 Rotavapor System
- Speed Range of 20 -280 rpm
- 50ml -4l Capacity
- Temperature Range from 20°C to 180°C
- Full 180 Day Parts and Labor Warranty
Go to Product Details
Buchi R-100 Rotary Vap System
Rotation Speed
20 to 280 rpm
- Brushless motor with high reliability.
- Internal capacity for up to 750 ml in tubes or 16 microplates.
- Accommodates various rotors including F-45-30-11, A-4-81, and A-4-62.
- Maximum speed of 14,000 RPM and a maximum g-force of 20,800 x g.
- Compact design with a height of only 13 inches, suitable for limited spaces.
- Dynamic temperature range from -9°C to 40°C.
- Onboard LCD display for rotation speed, time left in the cycle, and internal temperature.
The Buchi R-100 Rotary Vap System is used for single-staged distillations at high speeds with low stress. The R-100 includes "CR" glassware and the Buchi B-100 Heating Bath. The digital heating bath can be adjusted from 20° to 95°C and has a volume of 4L. Modular construction makes it possible to configure units with different glassware assemblies. With help from the V-100 Pump, the R-100 is capable of reaching low vacuum levels in minutes. The V-100 allows for a final vacuum of less than 10 mbar, and pumping volume of 1.5 meters cubed per hour. An onboard multifunctional valve unit allows for precise vacuum control along with irregular vacuum source containment. Modular construction allows for ultimate versatility, making it possible to configure units with a variety different glassware assemblies. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Buchi R-100 Rotary Vap System includes:
- > "C" Glassware w/ Dry Ice Trap
- > 1 Liter Evaporating Flask
- > 1 Liter Receiving Flask
- > Buchi B-100 Heating Bath
- > V-100 Diaphragm Pump
- > I-100 Interface Module
- > Vacuum Tubing and Connectors
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Buchi R-100 Rotary Vap System Specifications
Rotation Speed
|
20 to 280 rpm
|
Temperature
Control
|
20 ° to 95°C
|
Bath Capacity
|
4L
|
Pressure Load
on Condenser Coil
|
Max 6 bar
|
V-100 Vacuum Pump
|
Final Vacuum Possible
|
< 10 mbar
|
Pumping Volume
|
1.5mᶟ/h
|
Power Consumption
|
150 W
|
B-100 Heating Bath
|
Control
|
Analog Control Knobs
|
Temperature range
|
20 to 95
°C
|
Temperature Accuracy
|
+ 1°C
|
Buchi R-100 Rotary Vap Manual
Click Here to View the Technical Specs for the R-100
Buchi R-210 Rotavapor System
- Speed Range of 20 -280 rpm
- 50ml -4l Capacity
- Temperature Range from 20°C to 180°C
- Full 180 Day Parts and Labor Warranty
Go to Product Details
- Inspect all glassware for chips or cracks, reapplying vacuum grease to stopcocks to ensure a secure seal
- Examine displays, replacing any dim or dead pixels to maintain clear visibility
- Verify drive motor RPM using a tachometer for accurate rotational speed
- Confirm the ease of raising and lowering the tower, lubricating as needed for smooth operation
- Check the Woulff bottle for any leaks or structural breakages, ensuring airtight performance
- Disassemble and inspect the vacuum pump, cleaning and replacing seals or components as necessary
- Inspect internal wiring for each component, checking for any frays, cuts, or potential breakages
- Examine all hoses, hose barbs, and seals for signs of wear or leaks, ensuring integrity
- Inspect the vapor duct seal and replace if compromised to maintain vacuum efficiency
- Verify and calibrate the vacuum controller for accurate pressure regulation
- Confirm all hoses and cables are securely fastened to prevent disconnections or leaks
- Check heat bath temperatures against settings to ensure precise thermal control
- Confirm the final system vacuum level meets manufacturer specifications
- Ensure cosmetic condition meets standards, with no damage impacting function or appearance
Technician Inspection and Recertification Checklist: Rotary Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect all glassware for chips or cracks, apply vacuum grease to stopcocks
|
Completed
|
|
Examine displays and replace dim/dead pixels
|
Completed
|
|
Verify drive motor RPM with tachometer
|
Completed
|
|
Confirm ease of raising/lowering tower, lubricate if needed
|
Completed
|
|
Check Woulff bottle for leaks or structural issues
|
Completed
|
|
Disassemble and inspect vacuum pump, clean/replace parts as needed
|
Completed
|
|
Inspect internal wiring for frays, cuts, or breakages
|
Completed
|
|
Examine hoses, hose barbs, and seals for leaks
|
Completed
|
|
Inspect vapor duct seal, replace if compromised
|
Completed
|
|
Verify and calibrate vacuum controller
|
Completed
|
|
Confirm hoses and cables are securely fastened
|
Completed
|
|
Check heat bath temperatures for accurate control
|
Completed
|
|
Confirm final system vacuum level meets specifications
|
Completed
|
|
Ensure cosmetic condition meets standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|