Buchi B-490 Heating Bath
Temperature Range 20°C to 180°C
Max Speed ±1°C at 20°C to 100°C
- Corrosion-resistant PTFE coating for durability.
- Includes an insulated outer wall and two overtemperature shutoffs for added safety.
- Equipped with a digital LED display for real-time temperature monitoring and setpoint adjustments.
- Designed for use with R-200 and R-205 Rotavapor Systems.
- Analog for user-friendly operation.
- Maintains ±1°C accuracy at 20°C to 100°C and ±2°C at 100°C to 180°C.
The Buchi B-490 Heating Bath can accommodate evaporating
flasks ranging in size from 50 to 3000mL, with an overall holding capacity of 5L.
This heating bath can be adjusted from 20° to 180°C and has a temperature
accuracy of + 1ºC, making it suitable for high/low boiling solvents. The Buchi
B-490’s aluminum heating pan features a corrosion-resistant PTFE coating and,
for added safety, an insulated outer wall and two overtemperature shutoffs. The
waterbath features a digital LED display, which shows real-time temperature and
setpoint. (The B-490 is for use with and included in evaporating systems
R-200 and R-205.)
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Buchi B-490 Heating Bath includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Buchi B-490 Heating Bath Specifications
Temperature Range
|
20°C to 180°C
|
Temperature Accuracy
|
±1°C at 20°C to 100°C
±2°C at 100°C to 180°C
|
Control
|
Analog
|
Capacity
|
3L Flask
|
Compatibility
|
R-200 and R-205 Rotavapor Systems
|
Electrical Requirements
|
120V, 50/60Hz, 1400W
|
Display
|
Digital
|
Overall Dimensions
|
28 x 19.1 x 25 cm
|
Buchi B-490 Heating Bath Manual
Click Here to View the Technical Specs for the Buchi B-490 Heating Bath
Marshall Scientific Buchi B490 Heating Bath
- Confirm power supply and PCB are outputting stable, correct voltages across all essential circuits to ensure reliable operation
- Inspect the display for any dim, flickering, or dead pixels, replacing any compromised sections to maintain full clarity and accurate readouts
- Verify that all buttons, knobs, and switches are functioning properly, providing precise and responsive control across all settings
- Conduct a thorough inspection of the water reservoir for any holes, rust, or other structural damage that may impact its integrity
- Fill the reservoir with water, checking closely for any external or internal leaks that could affect performance
- Examine the lid for any signs of wear or damage, replacing hinges and handles if necessary to ensure a secure and proper seal
- Inspect for any internal leaks originating from the water reservoir, confirming the unit’s internal plumbing integrity
- Conduct a comprehensive inspection of internal wiring for any breakages, tears, or wear; replace any underperforming or damaged components to maintain safe operation
- Examine internal insulation for signs of burns, breaks, or deterioration, replacing as needed to ensure thermal efficiency and safety
- Confirm that all feet are intact and that the bath sits level and balanced on a flat surface to prevent movement or instability
- Calibrate both temperature accuracy and uniformity using a NIST-traceable temperature probe, verifying compliance with specification standards
- Perform a detailed cosmetic inspection to confirm that all external surfaces and finishes meet aesthetic and quality standards
Technician Inspection and Recertification Checklist: Water Bath
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm power supply and PCB output stable, correct voltages across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace for clarity
|
Completed
|
|
Verify functionality of all buttons, knobs, and switches for responsive control
|
Completed
|
|
Inspect water reservoir for holes, rust, or structural damage
|
Completed
|
|
Fill reservoir with water; check for external or internal leaks
|
Completed
|
|
Examine lid for wear or damage; replace hinges or handles as needed
|
Completed
|
|
Inspect for internal leaks from water reservoir to confirm plumbing integrity
|
Completed
|
|
Check internal wiring for damage; replace underperforming components for safety
|
Completed
|
|
Examine internal insulation for burns, breaks, or deterioration; replace as needed
|
Completed
|
|
Confirm feet are intact and bath is balanced on flat surface
|
Completed
|
|
Calibrate temperature accuracy and uniformity with NIST-traceable probe
|
Completed
|
|
Perform cosmetic inspection to confirm quality of external surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|