Boekel Scientific 133000 Economy Small Digital Incubator
The Boekel Scientific 133000 Economy Small Digital Incubator is designed to meet the needs of laboratories, hospitals, and pharmacies, offering an efficient and affordable solution for various medical and research procedures that require heating. This compact unit is equipped with a solid door and a small footprint, making it ideal for lab benchtops with limited space. With a volume of 0.8 cubic feet (22.7 liters), this incubator is a practical choice for everyday laboratory use.
It includes two shelves and can accommodate up to four shelves for versatile storage options. The incubator features a digital temperature readout and an analog control dial for precise temperature adjustments. Temperature settings range from Ambient +5°C to 60°C, with a temperature uniformity of ± 1°C at 37°C and stability of ± 1.5°C at the same temperature. Other notable features include a magnetic latch, an LCD temperature display, and a heating indicator light.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Boekel Scientific 133000 Economy Small Digital Incubator includes:
- > 2 Shelves
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Boekel Scientific 133000 Economy Small Digital Incubator Specifications
Temperature Range
|
Ambient +5°C to 60°C
|
Temperature Uniformity
|
± 1°C at 37°C
|
Temperature Stability
|
± 1.5°C at 37°C
|
Volume
|
0.8 cu. ft. (22.7L)
|
Shelf Area
|
1.7 ft.
2
|
Internal Dimensions (W x D x H)
|
11.5 x 12 x 10.5 in. (29.2 x 30.5 x 26.7 cm)
|
External Dimensions (W x D x H)
|
13 x 14 x 16.5 in. (33 x 35.6 x 41.3 cm)
|
Weight
|
22 lb. (10 kg)
|
Electrical Requirements
|
120V, 60Hz, 90W, 0.75A
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|