Biotek Synergy 2 SL Microplate Reader
The Biotek Synergy 2 SL is a single-channel microplate reader for research and
development and in vitro diagnostic use and includes incubation control, shaking, Gen5 software and Luminescence capability. Synergy 2 uses a unique combination of monochromator, filters and dichroic mirrors. Its three broad-spectrum light sources have been chosen for optimal illumination and excitation in all applications. The reader is computer-controlled using BioTek’s Gen5 PC software for all operations including data reduction and analysis. The Synergy 2 is robot accessible and compatible with BioTek’s BioStack Microplate Stacker. Gen5 supports OLE automation to facilitate the Synergy 2’s integration into an automated system. Luminescence is measured by the low-noise PMT detector through an empty filter position in the Emission filter wheel. Filters can also be used if light filtering is necessary. The Synergy 2 SL has a 4-Zone temperature control from 4°C over ambient to 50°C. Internal plate shaking is supported to ensure that reagents are properly mixed prior to reading. Please Note: This is the SL model of the Synergy 2 Microplate Reader, which is only configured for measuring/detecting Luminescence. Please see the Product Manual for more information.
The Biotek Synergy 2 SL Microplate Reader - Luminescence Only includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > Operator's Manual
- > Interface Cable
- > Full 180 Day Part and Labor Warranty
Biotek Synergy 2 SL Microplate Reader Specifications
Microplate Capacity
|
Standard 6-, 12-, 24-, 48-, 96-, 384-well microplates
Max Plate Height:
Luminescence modes: plates up to 1.25” (31.75 mm) high
|
Dimensions (D x W x H)
|
17.5 x 15 x 17.5 in
44.5 cm x 38.5 cm x 44.5 cm
|
Weight
|
57lbs (25.8 kg)
|
Environment
|
Operational temperature 18º to 40°C
|
Humidity
|
10% to 85% relative humidity (non-condensing)
|
Power Supply
|
24-volt external power supply compatible with
100-240 V~; +10% @50-60 Hz
|
Power Consumption
|
250W maximum
|
Incubation
|
Temperature control range from 4º over ambient to 50ºC.
Temperature variation ± 0.50°C across the plate @ 37°C (250 µl per
well with the plate sealed)
|
Plate Shaking
|
Low, Medium, High and Variable shaking speeds. Shake time is
programmable by the user
|
Luminescence
|
50 amol/well flash ATP in a 96-well plate (low-noise #49984 PMT)
500 amol/well flash ATP in a 96-well plate (red-shifted #49721 PMT)
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
BioTek Synergy 2 SL Manual Click Here to View the Technical Specs for the BioTek Synergy 2 SL Microplate Reader
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|