BioTek ELx405R Microplate Washer
Microplate Type 96-Well (Low Profile & Standard)
Manifold Type 96 Tube Dual Action
Wash Speed 96-wells, 300 µL/well, 3 cycles: <30 seconds
- Efficient and compact design, ideal for all laboratory settings
- Patented dual-action manifold for quick and thorough cleaning of 96-well plates
- User-friendly programming through a simple keypad interface with customizable control settings
- Features fully programmable fluid volumes, flow rates, shaking, and soaking options
- Includes AutoPrime and AutoRinse functions for easy maintenance and preparation
- Capable of handling both low profile and standard height 96-well microplates
- Offers selectable vacuum filtration with a range of -38 to -506 mmHg and time range from 5 to 999 seconds
- Dispense precision of ≤3% CV at 300 µL/well, with a residual volume of <2 µL/well in standard 96-well plates
The Biotek ELx405R microplate washers are both efficient and
space conscious in all lab settings. The BiotekELx405R platewasher features a
patented dual-action manifold to ensure fast operating time and overall
cleaning efficiency.
Programming is made easy through simple keypad interface with independent control of plate position during dispense and aspiration cycles, as well as fully programmable fluid volumes and flow rates, shaking, and soaking. Easy maintenance with a selection of pre-programmed routines for problem-free operation and run protection via fluid detection, fluid flow, and vacuum sensors. Among these programs, the ELx405R also employs unique AutoPrime and AutoRinse functions for both speedy maintenance
and quick prep times.
This Biotek ELx405 microplate washer is in excellent condition and is tested to
perform at factory specifications.
The BioTek ELx405R Microplate Washer includes:
- > (2) 4 Liter Waste Bottles
- > (1) 4 Liter Buffer Bottle
- > Vacuum Pump
- > Stainless Steel Bottle Holder
- > Microplate Carrier
- > Vacuum Tubing with Connections
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
BioTek ELx405R Microplate Washer Specifications
Microplate Types
|
96-well
Low profile and standard height
|
Onboard Software
|
Create, edit and run multiple protocols
|
Soaking
|
Programmable up to 60 minutes
|
Manifold Types
|
96 Tube Dual Action manifold for 96-well washing
|
Volume Range
|
50 to 3000 microliter increments
|
Wash Cycles
|
1 -10
|
Buffer/reagent Selection
|
Auto switching module for up to 4 buffers (option)
|
Supply bottle
|
4 L
|
Dispense Precision
|
≤3% CV: 300 µL/well (96-well washing)
|
Residual Volume
|
<2 µL/well (standard)
|
Wash Speed
|
96-wells, 300 µL/well, 3 cycles: <30 seconds
|
Flow Rates
|
High flow to low flow
Optimized rates for cell assays
|
Sterilization
|
Chemical
|
Vacuum Filtration
|
Selectable: Approximate final at 30 seconds:
Lowest: -38 mmHg
Low: -113 mmHg
Medium: -200 mmHg
High: -390 mmHg
Highest: -506 mmHg
Vacuum filtration time range: 5 to 999 seconds
|
Biotek ELX 405 Microplate Washer - Marshall Scientific
- Install new waste and buffer bottles to ensure proper fluid containment and system functionality.
- Integrate a new pump into the system, verifying compatibility and secure connections.
- Verify that the power supply and PCB output the correct voltages for reliable component operation.
- Conduct a self-test on the microplate washer, evaluating all operational parameters.
- Inspect the X-axis motor belt for signs of wear, such as tears, frays, or breakages, replacing if necessary.
- Examine dispense and aspirate tubes for any damage that could affect fluid flow accuracy.
- Confirm internal pump functionality, replacing seals as needed to maintain optimal performance.
- Inspect all internal wiring for frays, tears, or breakages, addressing any faults for electrical safety.
- Verify functionality of all internal and external sensors, ensuring accurate readings across the unit.
- Disinfect the unit using an appropriate disinfection cycle to ensure sterility and safety.
- Check that residual aspiration levels are within the manufacturer’s specified limits for accuracy.
- Conduct a final cosmetic inspection to verify that all external surfaces meet appearance standards.
Technician Inspection and Recertification Checklist: Microplate Washer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Install new waste and buffer bottles | Completed | |
Integrate new pump into system, verifying secure connections | Completed | |
Verify power supply and PCB output correct voltages | Completed | |
Conduct microplate washer self-test and evaluate results | Completed | |
Inspect X-axis motor belt for wear, replacing if necessary | Completed | |
Examine dispense and aspirate tubes for any damage | Completed | |
Confirm internal pump functionality, replace seals as needed | Completed | |
Inspect internal wiring for frays, tears, or breakages | Completed | |
Verify functionality of all internal and external sensors | Completed | |
Disinfect unit with appropriate disinfection cycle | Completed | |
Check residual aspiration levels meet specifications | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|