Biorad PowerPac Basic Electrophoresis Power Supply
- Programming in a compact, stackable case
- Timer control and constant voltage or constant current output
- Ability to pause and resume a run
- Automatic crossover featured to prevent damage
- When the set limit of the non-constant parameter is reached, and the power capability of the unit is not exceeded
- Used for nucleic acid gel electrophoresis & protein gel electrophoresis
The PowerPac Basic provides constant voltage or constant current to instruments used in
electrophoresis. The power supply operates at the values specified for the constant parameter. However, to prevent damage to the electrophoresis cell, the PowerPac Basic provides automatic crossover to constant current or constant voltage, depending on which set value is first reached. When the set limit of the non-constant parameter is reached, and the power capability of the unit is not exceeded, the power supply will switch, making the non-constant parameter the new constant parameter
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Biorad PowerPac Basic Electrophoresis Power Supply includes:
- > Power Cord
- > Calibration
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Biorad PowerPac Basic Electrophoresis Power Supply Specifications
Output
|
Voltage
|
10 V to 300 V, fully adjustable in 1 V steps
|
Voltage Accuracy
|
+/- 2% of reading or 3 volts, whichever is more
|
Current
|
4 mA to 400 mA, fully adjustable in 1 mA steps
|
Current Accuracy
|
+/- 2% of reading or 3 mA, whichever is more
|
Power (maximum)
|
75 Watts
|
Terminals
|
4-pair recessed banana jacks, floating in parallel
|
Timer Control
|
001 to 999 minutes, fully adjustable
|
Ripple
|
<1% @ 300 V and 70 W
|
Line Regulation
|
<1% @ 300 V and 70 W, 90-132 or 198-264 VAC
|
Load Regulation
|
<1% @ 300 V for a 50% change in output load
|
Drift
|
<1% @ after 15 min warmup at 300 V and 70 W
|
Noise
|
55 dBA at 1 meter
|
BioRad PowerPac Basic Manual
Click Here to View all the Technical Specs for the Biorad PowerPac Basic
- Verify internal power supply output for correct voltage in alignment with equipment specifications
- Inspect LED display for any dim or dead pixels, ensuring optimal readability
- Examine internal wiring and connections, checking for any rips, tears, or compromised insulation
- Ensure fan is clean and fully operational to maintain adequate cooling
- Test PCB test points for proper voltage output, confirming consistent circuit performance
- Validate functionality of all buttons and switches, ensuring responsive control
- Measure and confirm constant voltage, amplitude, and wattage using precise resistance values
- Cross-check displayed voltage with multimeter readings for accuracy
- Monitor for rapid resistance changes, software errors, and power fluctuations
- Inspect line cord for any signs of wear, cuts, or damage impacting safety
- Confirm cosmetic condition meets quality standards, free of any damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Electrophoresis Power Supplies
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify internal power supply output voltage
|
Completed
|
|
Inspect LED display for dim or dead pixels
|
Completed
|
|
Examine internal wiring for rips or tears
|
Completed
|
|
Ensure fan is clean and fully operational
|
Completed
|
|
Test PCB test points for voltage accuracy
|
Completed
|
|
Validate functionality of buttons and switches
|
Completed
|
|
Confirm constant voltage, amplitude, and wattage with resistance
|
Completed
|
|
Cross-check displayed voltage with multimeter
|
Completed
|
|
Monitor for rapid resistance changes and power fluctuations
|
Completed
|
|
Inspect line cord for wear or damage
|
Completed
|
|
Confirm cosmetic condition meets quality standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|