Popular Alternative Choice
Accuris E2502-400 PowerVolt Power Supply
- Adjustable output: 10 to 500 VDC & 4 to 400mA with a max power of 120W.
- Features capacitive touch screen control with color TFT LCD display.
- Timer function from 1 min to 20 hours, or use in continuous mode.
- Compact & portable: 3.5 x 5 x 8.5 in., weighing 2.2 lb, and operable 100 to 240VAC.
Go to Product Details
Free Shipping!
Biorad PowerPac 1000 Electrophoresis Power Supply
- Provides constant voltage, power, and current for a variety of electrophoresis applications
- Ideal for isoelectrofocusing (IEF), blotting, general electrophoresis, SDS-PAGE, and more
- Adjustable voltage output from 5 to 1000 volts dc in 1-volt increments
- Current output can be adjusted from 1 to 500 milliAmps in 1 mA increments
- Power output is adjustable from 1 to 250 watts in 1 W increments
- Fully programmable, allowing storage of up to nine separate methods with up to nine steps each
The Biorad PowerPac 1000 Electrophresis Power Supply provides constant voltage, power, and current, ideally for isoelectrofocusing (IEF) applications, blotting, and general electrophoresis, as well as SDS-PAGE, two-dimensional electrophoresis, native gel electrophoresis, and horizontal DNA/RNA electrophoresis. The voltage output is adjustable from 5 to 1000 volts dc in increments of 1 volt. The current output is adjustable from 1 to 500 milliAmps in increments of 1 mA. The power output is adjustable from 1 to 250 watts in increments of 1 W. Fully programmable, the PowerPac 1000 can store up to nine separate methods with up to nine steps each.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Biorad PowerPac 1000 Electrophoresis Power Supply includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Biorad PowerPac 1000 Electrophoresis Power Supply Specifications
Biorad PowerPac 1000 Power Supply
|
Max Voltage: 1000V
|
Max Current: 500mA
|
Max Wattage: 250W
|
Biorad Powerpac 1000 Manual
Click Here to View all the Technical Specs for the PowerPac 1000
- Verify internal power supply output for correct voltage in alignment with equipment specifications
- Inspect LED display for any dim or dead pixels, ensuring optimal readability
- Examine internal wiring and connections, checking for any rips, tears, or compromised insulation
- Ensure fan is clean and fully operational to maintain adequate cooling
- Test PCB test points for proper voltage output, confirming consistent circuit performance
- Validate functionality of all buttons and switches, ensuring responsive control
- Measure and confirm constant voltage, amplitude, and wattage using precise resistance values
- Cross-check displayed voltage with multimeter readings for accuracy
- Monitor for rapid resistance changes, software errors, and power fluctuations
- Inspect line cord for any signs of wear, cuts, or damage impacting safety
- Confirm cosmetic condition meets quality standards, free of any damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Electrophoresis Power Supplies
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify internal power supply output voltage
|
Completed
|
|
Inspect LED display for dim or dead pixels
|
Completed
|
|
Examine internal wiring for rips or tears
|
Completed
|
|
Ensure fan is clean and fully operational
|
Completed
|
|
Test PCB test points for voltage accuracy
|
Completed
|
|
Validate functionality of buttons and switches
|
Completed
|
|
Confirm constant voltage, amplitude, and wattage with resistance
|
Completed
|
|
Cross-check displayed voltage with multimeter
|
Completed
|
|
Monitor for rapid resistance changes and power fluctuations
|
Completed
|
|
Inspect line cord for wear or damage
|
Completed
|
|
Confirm cosmetic condition meets quality standards
|
Completed
|
|