Popular Alternative Choice
Scilogex TC1000-S Thermal Cycler
- High-performance with color LCD touch screen for easy programming.
- Excellent temp uniformity, ideal for 96 x 0.2mL PCR tubes or plates.
- Temperature range of 4 - 99C, with a heating/cooling rate of 5.0C/min and 0.5C/min respectively.
- Compact design: 15"L x 11"W x 10"H, weighing 25lbs and operable from 110-240V.
Go to Product Details
No Lead Time and Free Shipping!
Biorad MyCycler Thermal Cycler
Temperature Range 4°C to 100°C
Ramping Rates Heating: 2.5°C/sec - Cooling: 1.5°/sec
- Compact 96-well PCR system with gradient feature for gene amplification
- Intuitive programming with templates for easy protocol setup and editing
- Capable of programming up to 9 cycles with 9 steps each, repeating up to 99 times per cycle
- Self-adjusting heated lid to maintain accurate temperature for standard 0.2 ml thin-wall PCR tubes, strips, and plates
- Offers temperature monitoring and control via algorithm or block modes
- Rapid cycling speed with a maximum heating rate of at least 2.5°C/sec and cooling rate of 1.5°C/sec
- Temperature range from 4-100°C with well-to-well uniformity of ±0.5°C
- Functional temperature gradient ranging from 5-25°C, with a gradient accuracy of ±0.5°C
The Biorad MyCycler Thermal Cycler, 115V, is a sleek 96-well PCR system with a gradient feature used for gene amplification. This compact thermal cycler features intuitive programming of temperatures, times, and protocols with ready-made templates for ease of viewing, editing, and running protocols. Program protocols with as many as 9 cycles with 9 steps each and repeats up to 99 times per cycle. The MyCycler’s self-adjusting heated lid maintains accurate temperature and accommodates standard 0.2 ml thin-wall PCR tubes, strips, and plates. Temperature monitoring and control can be performed either via algorithm, which uses the experimentally derived sample temperature given specific sample volume, or block, which uses the actual sample block temperature. The Biorad MyCycler employs multiple built-in utilities, such as self-diagnostics and storage of detailed run and validation reports, for optimal performance for PCR and other thermal cycling techniques.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Biorad MyCycler Thermal Cycler includes:
- > 96 Well Alpha Block
- > Power Cord
- > Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Biorad MyCycler Thermal Cycler Specifications
Block Temperatures
|
Range
|
4-100°C
|
Cycling Speed
|
With 20 ul sample in a 0.2 ml tube, after warm up, when cycling
between 50°C and 95°C; at least 2.5°C/sec maximum heating rate and 1.5°C/sec
maximum cooling rate
|
Storage Cooling
|
Less than 3 minutes to cool from 40°C to 4°C
|
Overshoot
|
Less than0.5°C
|
Accuracy
|
±0.5°C to NIST traceable standard after 30 seconds at 90°C, to block
mode
|
Uniformity
|
Well to well uniformity of ±0.5°C within 30 seconds after arrival
at 90°C
|
Temperature Gradient
|
Input Gradient Range
|
1-25°C
|
Functional gradient range
|
5-25°C
|
Mininum temperature of gradient
|
30°C
|
Maximum temperature of gradient
|
99°C
|
Gradient accuracy
|
±0.5°C
|
Gradient Uniformity
|
0.5°C
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|