Popular Alternative Choice
Scilogex TC1000-S Thermal Cycler
- Scilogex TC1000-S: High performance and throughput via color LCD touch screen.
- Capable of handling 96 x 0.2mL PCR tubes or 96-well plates with precise temperature control.
- Temperature range of 4 - 99C with rapid heating/cooling rates for efficient operation.
- Included: 96 Well Block, power cord, guides, and a 1-year manufacturer warranty.
Go to Product Details
No Lead Time and Free Shipping!
Biorad iCycler Thermal Cycler
The Biorad iCycler is a 96-well gradient thermal cycler
equipped to handle everyday demands of PCR applications with extremely
reproducible quantitation and excellent uniformity. This thermal cycler features
built-in PCR protocols and an easy-to-use interface. The iCycler’s gradient can
accommodate 2ml and 5ml tubes, strips, and plates, allowing for increased
cycling speed and greater throughput.
The Biorad iCycler can be used for clinical diagnostics, gene
expression and amplification, and measuring viral loads and pathogen detection,
among other applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Biorad iCycler Thermal Cycler includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Biorad iCycler Thermal Cycler Specifications
Temperature
|
Range
|
4° to 100°C
|
Cycling Speed
|
96 x 0.2 ml Reaction Module: with 20 µl sample in a 0.2 ml tube in
the 96 well module, after full warm up, when cycling between 50°C and 95°C;
at least 3.3°C/sec maximum heating rate and 2.0°C/sec maximum cooling rate.
|
Storage Cooling
|
less than 3 minutes to cool from 40°C to 4°C
|
Overshoot
|
less than 0.5°C
|
Accuracy
|
±0.3°C to NIST traceable standard after 30 seconds at 90°C,
block mode
|
Uniformity
|
well-to-well uniformity of ±0.4°C within 30 seconds after
arrival at 90°C
|
Temperature Gradient
|
Input Gradient Range
|
1° to 25°C
|
Functional Gradient Range
|
5° to 25°C
|
Min Temp Gradient
|
40°C
|
Max Temp Gradient
|
99°C
|
Accuracy
|
±0.4°C
|
Uniformity
|
±0.4°C
|
Programming
|
Number of Programs
|
256 on board
|
Cycles/Protocol
|
up to 9
|
Steps/Cycle
|
up to 9
|
Repeats/Cycle
|
up to 600
|
Dwell Time
|
00:01 – 99:59 minutes
|
Pause
|
Yes
|
Built-In Protocols
|
Templates for cycle sequencing, Touchdown, RT PCR, nested PCR, long
PCR
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|