Popular Alternative Choice
Biorad C1000 Thermal Cycler
- Biorad C1000: Large color display, graphical/text programming & windows OS.
- Storage & Connectivity: Use USB Flash Drive, connect up to 3 cyclers or 32 with PC.
- Interchangeable module: Comes with a 96-well fast reaction module for flexibility.
- Includes: 96 Well Block, Calibration Report, Power Cord, & 180 Day Parts and Labor Warranty.
Go to Product Details
Fully Recertified!
Biorad C1000 Touch Thermal Cycler
Gradient Range 30 °C to 100 °C
Temperature Range 0 °C to 100 °C
- Superior Performance: The C1000 Touch thermal cycler offers excellent performance for PCR processes.
- Large color touch screen makes programming straightforward and user-friendly.
- Fully modular platform accommodates varying throughput needs with interchangeable reaction modules that can be swapped quickly and without tools.
- Each module features a heated lid that is adjustable for various vessels and sealers, including both low-profile and standard-height PCR plates.
- Features two programming options, including the Protocol Autowriter which automatically generates customized standard, fast, and ultrafast protocols.
- Increase throughput capability by linking up to 3 additional S1000 thermal cyclers or adding PC control for up to 32 cyclers.
- Quickly and easily identify optimal annealing temperatures by using the programmable temperature gradient
- Ideal for applications and uses with amplification/PCR, cloning, cycle sequencing, gene expression studies, and mutagenesis.
- The C1000 Touch can be used with six different reaction modules for flexibility.
The Biorad C1000 Touch Thermal Cycler provides superb thermal performance for fast, reliable results. This fully modular platform supports interchangeable reaction modules, including two optical modules for real-time PCR that swap in seconds without requiring tools. Equipped with a large state-of-the-art touch screen, the C1000 Touch Thermal Cycler offers a choice of programming methods, including graphical and automatic. The C1000 Touch is compatible with any 1000-series reaction module, with each including a fully adjustable heated lid that is capable of running reliably with a broad range of reaction vessels. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Biorad C1000 Touch Thermal Cycler includes:
- > 96 Well Block
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Biorad C1000 Touch Thermal Cycler Specifications
Input Power
|
Up to 850W, maximum
|
Frequency
|
50-60 Hz, single phase
|
Display
|
8.5 inch LCD display and touch screen
|
Ports
|
5 USB A, 1 USB B
|
Memory
|
>1,000 typical programs onboard, unlimited with USB flash drive
|
Dimensions (W x D x H)
|
13 x 18 x 8 in. (33 x 46 x 20 cm)
|
Weight
|
23 lb. (10 kg.)
|
Temperature Control Modes
|
Calculated and block
|
PCR License
|
Yes
|
Programing Options
|
Step-based graphical and automatic
|
Reporting
|
Exportable run logs, system error logs
|
Onboard Software
|
Windows CE 6.0
|
PC Compatibility
|
Windows XP or higher
|
USB Peripheral Compatibility
|
Mouse, USB flash drive, bar code reader
|
Real-time PCR upgrade
|
6-channel, 5-color CFX96 or 5-channel, 4-color CFX384 optical reaction module (not included)
|
Instant Incubation
|
Yes
|
Security Features
|
Optional log-in required mode for regulated environments
|
Reaction Module: 96 Well Fast Specifications
Sample Capacity
|
96 x 0.2 mi tubes
|
Maximum Ramp Rate
|
5°C/sec
|
Average Ramp Rate
|
3.3°C/sec
|
Temperature Range
|
0-100°C
|
Temperature Accuracy
|
±0.2° of Programed target at 90°C
|
Temperature Uniformity
|
±0.4°C well to well within 10 sec of arrival at 90°C
|
Gradient Capability
|
Yes
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|