Biorad C1000 Thermal Cycler w/ Dual 48 Well Block
Sample Capacity 2 x 48 x 0.2 ml tubes
- Compatible with 1000-series reaction modules, includes an interchangeable dual 48/48 well fast reaction module.
- Large color display supports intuitive graphical and text-based programming via a Windows OS.
- Enhanced throughput with options to control up to 32 cyclers, facilitated by USB flash drive data management.
- High-resolution 12 x 9 cm display with multiple USB ports for expansive connectivity.
- Stores over 1,000 programs, supporting extensive PCR protocol customization.
- Dual 48-well block features precise temperature control with calculated and block modes.
- Rapid temperature ramp rates of up to 4.0°C/sec, with an average of 3.0°C/sec for efficient sample processing.
- Temperature accuracy of ±0.2° at 90°C and uniformity of ±0.4°C across wells ensures consistent PCR results.
- Gradient capability for optimizing protocols through variable temperature settings.
The Biorad C1000 Thermal Cycler with Dual 48 Well Block is engineered for high-throughput PCR applications, featuring an interchangeable dual 48/48 well fast reaction module that integrates seamlessly with any 1000-series reaction modules. This device boasts a large, high-resolution 12 x 9 cm color display for intuitive graphical and text-based programming, enhanced by Windows CE 6.0 OS. It supports the storage and easy transfer of data via USB flash drive and can control up to 32 cyclers, increasing laboratory productivity. With precise temperature control modes, a maximum ramp rate of 4.0°C/sec, and superior temperature uniformity of ±0.4°C, this thermal cycler delivers exceptional performance and accuracy for critical scientific research.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Biorad C1000 Thermal Cycler w/ Dual 48 Well Block includes:
- > Dual 48 Well Block
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Biorad C1000 Thermal Cycler w/ Dual 48 Well Block Specifications
Input Power
|
400W, maximum
|
Frequency
|
50-60 Hz, single phase
|
Display
|
12 x 9 cm, high resolution
|
Ports
|
5USB A, 1 USB B
|
Memory
|
>1,000 Typical Programs
|
Dimensions (W x D x H)
|
13x 18 x 8
|
Weight
|
23 lb
|
Temperature Control Modes
|
Calculated and block
|
PCR License
|
Yes
|
Programing Options
|
Step-based graphical, text based, automatic
|
Reporting
|
Exportable run logs, system error logs
|
Onboard Software
|
Windows CE 6.0
|
PC Compatibility
|
Windows XP or higher
|
USB Peripheral Compatibility
|
Mouse, USB flash drive, bar code reader
|
Real-time PCR upgrade
|
6- Channel, 5-color CFX(^ module
|
Instant Incubation
|
Yes
|
Fuses
|
Two 6.3 A, 250 V x 220 mm
|
Reaction Module: Dual 48/48 Well Fast
Sample Capacity
|
2 x 48 x 0.2 ml tubes
|
Maximum Ramp Rate
|
4.0°C/sec
|
Average Ramp Rate
|
3.0°C/sec
|
Temperature Range
|
0-100°C
|
Temperature Accuracy
|
±0.2° of Programed target at 90°C
|
Temperature Uniformity
|
±0.4°C well to well within 10 sec of arrival at 90°C
|
Gradient Capability
|
Yes
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|