BioTek Synergy HTX Microplate Reader - Absorbance Only
Wavelength Range 200 to 999 nm
Wavelength Accuracy ± 2 nm
Wavelength Repeatability
± 0.2 nm
- Performs accurate absorbance measurements in various well formats.
- Offers continuous wavelength selection from 200 to 999 nm in 1-nm increments.
- Supports endpoint, kinetic, multiwavelength, and spectral scanning methods.
- Xenon flash lamp facilitates both UV and visible light absorbance measurements.
- Linear and orbital shaking modes with adjustable intensities, programmable by the user.
- Ideal for ELISA assays, kinetic enzyme assays, nucleic acid quantitation, protein analysis, and more.
The BioTek Synergy HTX
is a single-channel, automated, benchtop, absorbance microplate reader that performs optical density measurements of samples in a 6-, 12-, 24-, 48-, 96- or 384-well microplate formats. The Synergy HTX features a monochromator which offers continuous wavelength selection from 200 to 999 nm in 1-nm increments, as well as wavelength scanning without the need for interference filters. A variety of read methods can be utilized, including endpoint, kinetic, multiwavelength, and spectral scanning. This microplate reader also offers superior absorbance optical specifications, with an extended dynamic range of up to 4.000 OD. The user may choose from two reading speeds: normal and sweep mode. The microplate reader is computer controlled using BioTek's Gen5 PC software for all operations, including data reduction and analysis.
Absorbance measurements are made by utilizing the featured xenon flash lamp and the monochromator for wavelength selection. The use of a xenon flash lamp allows for both UV and visible light absorbance measurements. The Synergy HTX is also equipped a unique 4-Zone temperature control system, allowing for control from 4°C over ambient to 50°C that ensures superior temperature uniformity necessary for kinetic assays. An internal microplate shaker is also featured on this unit, which allows for selection between linear and orbital shaking modes with adjustable intensities all programmable by the user. The Synergy HTX can also be interfaced with BioTek's BioStack Microplate Stacker for robotic functionality. Common applications this microplate reader is used for includes: ELISA assays, kinetic enzyme assays, nucleic acid quantitation, protein analysis, and much more. Please Note: This is the S1A model of the Synergy HTX Microplate Reader, which is only configured for measuring/detecting Uv-Vis (Absorbance). Please see the Technical Specification Sheet/Product Manual for more information.
The BioTek Synergy HTX Microplate Reader - Absorbance Only includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > ABS and Self Test Calibration Report
- > Operator's Manual
- > Interface Cable
- > Full 180 Day Parts and Labor Warranty
BioTek Synergy HTX Microplate Reader - Absorbance Only Specifications
Wavelength Range
|
200 to 999 nm
|
Wavelength Accuracy
|
± 2 nm
|
Wavelength
Repeatability
|
± 0.2 nm
|
Wavelength Bandpass
|
2.4 nm
|
Microplate Capacity
|
Standard 6-, 12-, 24-, 48-, 96- and
384-well microplates
|
Dimensions (W x D x
H)
|
16 x 15 x 10 in (40.6 x 38 x 25.4 cm)
|
Weight
|
40 lb. (18 kg)
|
Environment
|
Operational
temperature 18º to 40°C
|
Light Source
|
Xenon
flash light source: 10W maximum average power
|
Humidity
|
10% to 85% relative
humidity (non-condensing)
|
Power Consumption
|
100 VA max
|
Incubation
|
Temperature control
range from Ambient +4º to 50ºC
Temperature
variation ± 0.5°C at 37°C
|
Plate Shaking
|
Linear: 1mm to 6mm amplitude, 18Hz to 6Hz frequency
Orbital Slow: 1mm to 6mm amplitude, 10Hz to 3Hz frequency
Orbital Fast: 1mm to 6mm amplitude, 14Hz to 5Hz frequency
|
Read Modes
|
Normal and
Sweep
|
Absorbance
Measurement Range
|
0.000 to 4.000 OD
|
Absorbance Accuracy
|
96-Well Plate, Normal Read Speed: 0.000 to 2.000 OD ± 1.0% ± 0.010 OD 2.000 to 3.000 OD ± 3.0% ± 0.010 OD
384-Well Plate, Normal Read Speed:
0.000 to 2.000 OD ± 2.0% ± 0.010 OD 2.000 to 2.500 OD ± 3.0% ± 0.010 OD
96-Well/384-Well Plate, Sweep Read Speed: 0.000 to 1.000 OD ± 1% ± 0.010 OD
|
Absorbance Linearity
|
96-Well Plate, Normal Read Speed: 0.000 to 2.000 OD ± 1.0% ± 0.010 OD 2.000 to 3.000 OD ± 3.0% ± 0.010 OD
384-Well Plate, Normal Read Speed:
0.000 to 2.000 OD ± 2.0% ± 0.010 OD 2.000 to 2.500 OD ± 3.0% ± 0.010 OD
96-Well/384-Well Plate, Sweep Read Speed: 0.000 to 1.000 OD ± 1% ± 0.010 OD
|
Absorbance
Repeatability
|
96-Well/384-Well Plate, Normal Read Speed 0.000 to 2.000 OD ± 1.0% ± 0.005 OD 2.000 to 3.000 OD ± 3.0% ± 0.005 OD
96-Well/384-Well Plate, Sweep Read Speed: 0.000 to 1.000 OD ± 2.0% ± 0.010 OD
|
Electrical Requirements
|
100-240V, 50/60Hz
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
Frequently Purchased With
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|