BioTek ELx808 Microplate Reader - Absorbance Only
Wavelength Range 340 to 900 nm
Microplate Format Standard 96-Well
Read Modes Endpoint, Kinetic, Linear Scan
- Eight-channel absorbance microplate reader covering 340 to 900 nm wavelength range
- Standard filters included at 340, 405, 450, 490, and 630 nm
- Extended dynamic range up to 4.000 OD with high optical precision
- Onboard processor with LED screen and membrane keys for protocol management
- Equipped with Gen5 Data Analysis Software for comprehensive control and data analysis
- 4-Zone Temperature Control up to 50°C for temperature-sensitive assays
- Internal microplate shaker with adjustable low, medium, high, and variable speeds
- Supports endpoint, kinetic, and linear scan read modes for flexible applications
- Accurate absorbance measurement with ±1.0% at 405 nm and ±1.0% at 340 nm
- Compact design with a tungsten halogen light source, operating in diverse voltage ranges
The BioTek ELx808 is an eight-channel, absorbance microplate reader-assay system that can measure the optical density of solutions in 96-well microplates between 340 nm and 900 nm. This unit operates using a filter based wavelength selection system, with 340, 405, 450, 490, and 630 nm filters included standard. The ELx808 features superior optical specifications, with an extended dynamic range of up to 4.000 OD. This microplate reader also features an onboard processor, 2 x 24 character LED screen, and membrane keys to allow for easy definition and management of assay protocols, templates, formulas, and data. Through the control of the included Gen5 Data Analysis Software however, the ELx808 becomes an efficient, easy to use and powerful system for a variety of microplate based applications.
The ELx808 Microplate Reader is capable of conducting read endpoint (regular), kinetic, as well as linear scan read modes. The ELx808 is also equipped a unique 4-Zone Temperature Control system, allowing for control from up to 50°C that provides excellent stability for temperature sensitive assays and long-term bacteria/yeast growth studies. An internal microplate shaker is also featured on this unit, which allows for selection between low, medium, high, and variable speeds which are all programmable by the user. Please Note: This is the IU model of the Elx808 Microplate Readers, which features the incubation to 50°C and 340 to 900 nm wavelength range. This Microplate Reader is only configured for measuring/detecting Uv-Vis (Absorbance).
The BioTek ELx808 Microplate Reader - Absorbance Only includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > ABS and Self Test Calibration Report
- > Operator's Manual
- > Interface Cable
- > Full 180 Day Parts and Labor Warranty
BioTek ELx808 Microplate Reader - Absorbance Only Specifications
Wavelength Range
|
340 to 900 nm
|
Filters Included
|
340, 405, 450, 490, and 630 nm
|
Wavelength Bandpass
|
10 nm
|
Microplate Capacity
|
Standard 96-well microplates
|
Dimensions (W x D x H)
|
16 x 15.5 x 8.75 in (40.6 x 39.4 x 22.2 cm)
|
Weight
|
30 lb. (13.6 kg)
|
Environment
|
Operating temperature 18ºC to 40°C
|
Light Source
|
Tungsten halogen-filled bulb
|
Humidity
|
10% to 85% relative humidity (non-condensing)
|
Power Consumption
|
100 VA, four voltage ranges accommodated by the voltage selection switch:
Range 1: 100 VAC, 90 to 110 VAC, 50/60Hz Range 2: 120 VAC, 108 to 132 VAC, 50/60Hz Range 3: 230 VAC, 207 to 253 VAC, 50/60Hz Range 4: 240 VAC, 216 to 264 VAC, 50/60Hz
|
Incubation
|
Temperature control up to 50ºC
Temperature uniformity of ± 0.5°C at 37°C
|
Plate Shaking
|
Low, Medium, High, and Variable Speeds
|
Read Modes
|
Endpoint, Kinetic, and Linear Scan
|
Read Speeds
|
Normal Mode: Single/Dual Wavelength 12/20 sec.
Rapid Mode: Single/Dual Wavenlength 8/13 sec.
Kinetic Interval Rapid Mode: Single Wavelength 6 sec
|
Absorbance Measurement Range
|
0.000 to 4.000 OD (400 to 900 nm range)
0.000 to 3.000 OD (340 to 400 nm range)
|
Absorbance Accuracy
|
96-Well Plate, Endpoint:
± 1.0% ± 0.010 OD from 0.000 to 2.500 OD @ 405 nm
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 340 nm
|
Absorbance Linearity
|
96-Well Plate, Endpoint:
± 1.0% from 0.000 to 2.500 OD at 405 nm
± 2.0% from 2.500 OD to 3.500 OD @ 405 nm
± 1.0% from 0.000 to 2.000 OD @ 340 nm
|
Absorbance Repeatability
|
96-Well Plate, Endpoint:
± 0.5% ± 0.005 OD from 0.000 to 2.500 OD @ 405 nm
± 1.5% ± 0.005 OD from 2.500 to 3.500 OD @ 405 nm
± 2.5% ± 0.005 OD from 3.500 to 4.000 OD @ 405 nm
± 1.0% ± 0.005 OD from 0.000 to 2.000 OD @ 340 nm
|
Electrical Requirements
|
100-240V, 50/60Hz
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|