BioTek ELx800 Absorbance Microplate Reader
Wavelength Range
400 to 750 nm (Filter-Based)
Filters Included 405, 450, 490, and 630 nm
Microplate Capacity
Standard 6-, 12-, 24-, 48-, and 96-well microplates
- Compact absorbance-only microplate reader, ideal for clinical and life science labs
- Supports 6- to 96-well microplates with end point, kinetic, and well area scanning modes
- Filter-based wavelength range of 400 to 750 nm, including 405, 450, 490, and 630 nm filters
- Fast reading speeds: 50 seconds in normal mode, 30 seconds in rapid mode for 96-well plates
- Accurate absorbance measurements from 0.000 to 3.000 OD, with a 10 nm bandpass
- Durable tungsten halogen lamp light source for reliable performance
- Operates within environmental conditions of 18°C to 40°C and 10% to 85% humidity
The BioTek ELx800 Microplate Reader is an indispensable tool for clinical and life science research laboratories, designed to efficiently handle absorbance-only readings. This robust, compact unit seamlessly integrates with BioTek’s Gen5 software, expanding its capabilities to include end point, kinetic, and well area scanning for a variety of microplate formats, from 6-well up to 96-well. Ideal for settings where space or budget might be restricted, the ELx800 boasts comprehensive on-board software that includes numerous curve fitting options, data transformations, and tools for assay validation. Its reliable performance is complemented by a filter-based wavelength range of 400 to 750 nm, including specific filters at 405, 450, 490, and 630 nm, and a 10 nm bandpass. Although lacking incubation and plate shaking features, the ELx800 excels in reading modes like endpoint, kinetic, and well area scanning, providing absorbance measurements from 0.000 to 3.000 OD with high precision. Reading speeds are swift, clocking at 50 seconds for normal and 30 seconds for rapid modes in 96-well plates, ensuring efficient throughput for busy labs. The combination of durability, versatility, and precision makes the ELx800 a superb value for enhancing laboratory productivity. Please Note: This is the standard model of the Elx800 Microplate Readers, which does not include a 340 nm filter and does not feature compatibility with 384-well plates. This Microplate Reader is only configured for measuring/detecting Uv-Vis (Absorbance).
All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioTek ELx800 Microplate Reader - Absorbance Only includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > 405, 450, 490, and 630 nm Filters
- > ABS and Self Test Calibration Report
- > Operator's Manual
- > Interface Cable
- > Full 180 Day Parts and Labor Warranty
BioTek ELx800 Microplate Reader - Absorbance Only Specifications
Wavelength Range
|
400 to 750 nm (Filter-Based)
|
Filters Included
|
405, 450, 490, and 630 nm
|
Wavelength Bandpass
|
10 nm
|
Microplate Capacity
|
Standard 6-, 12-, 24-, 48-, and 96-well microplates
|
Light Source
|
Tungsten halogen lamp
|
Incubation
|
N/A
|
Plate Shaking
|
N/A
|
Read Modes
|
Endpoint, Kinetic, and Well Area Scanning
|
Reading Speeds
|
Normal Mode: 50 sec. (96-Well)
Rapid Mode: 30 sec. (96-Well)
|
Absorbance Measurement Range
|
0.000 to 3.000 OD
|
Absorbance Accuracy
|
96-Well Plate, Normal Read Mode:
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD at 405 nm
96-Well Plate, Rapid Read Mode:
± 2.0% ± 0.020 OD from 0.000 to 2.000 OD at 405 nm
|
Absorbance Linearity
|
96-Well Plate, Normal Read Mode:
± 1.0% from 0.000 to 2.000 OD at 405 nm
± 3.0% from 2.000 OD to 3.000 OD at 405 nm
96-Well Plate, Rapid Read Mode:
± 2.0% from 0.000 to 2.000 OD at 405 nm
|
Absorbance Repeatability
|
96-Well Plate, Normal Read Mode:
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD at 405 nm
96-Well Plate, Rapid Read Mode:
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD at 405 nm
|
Onboard Software
|
55 user-definable programs
|
Environmental Operating Temperature
|
18C to 40C
|
Humidity
|
10% to 85% noncondensing
|
Dimensions (D x W x H)
|
16.5 x 15 x 7 in. (42 x 38 x 18 cm)
|
Weight
|
18.5 lb. (8 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
Frequently Purchased With
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|