BioTek 405 TSRVS Microplate Washer
The BioTek 405 TSRVS Microplate Washer is an advanced laboratory instrument designed for high-efficiency washing of 96-well microplates. It features a user-friendly high-resolution 5.7-inch touch screen with a color graphical display, powered by an Intel Atom processor running Windows CE. This intuitive interface makes it simple to develop protocols, perform maintenance, and operate the instrument daily. The washer is compatible with both low profile and standard height 96-well plates and uses a single 96-tube (8x12) manifold exclusively for 96-well washing. One of the key attributes of the 405 TSRVS is its versatile volume range, allowing for dispensing from 50 to 3,000 µL per well with a dispense precision of ≤ 3% CV at 300 µL/well. It also boasts a remarkably low residual volume of ≤ 2 µL/well in solid bottom plates. The washer is capable of automated internal switching between up to four wash buffers, accommodating a variety of assay requirements. Additionally, it offers a broad range of flow rates from 1 to 11, catering to everything from slow to fast, including specialized low flow rates for cell wash applications.
The BioTek 405 TSRVS excels in flexibility with its programmable shaking feature, allowing for durations up to 60 minutes and intensities ranging from slow to fast. The soak time is equally adaptable, programmable up to 60 minutes. Its efficiency is highlighted in its wash speed, capable of completing three cycles with 300 µL per well in a 96-well format in under 30 seconds. This model is particularly well-suited for cell-based assays, microsphere-based assays, and ELISA applications, thanks to its specialized features like 4-wash buffer switching, the ultrasonic advantage, and automated magnetic bead assay capabilities. The BioTek 405 TSRVS Microplate Washer is a sophisticated solution for laboratories seeking a reliable and versatile instrument for microplate washing.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioTek 405 TSRVS Microplate Washer includes:
- > (2) 4 Liter Waste Bottles
- > (4) 4 Liter Buffer Bottles
- > Vacuum Pump
- > Stainless Steel Bottle Holders
- > Microplate Carrier
- > Vacuum Tubing with Connections
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
BioTek 405 TSRVS Microplate Washer Specifications
Microplate Types
|
96-well
Low profile and standard height
|
Manifold Type
|
Single 96-tube (8x12) for 96-well washing only
|
Volume Range
|
50 to 3,000 µL/well
|
Buffer Selection
|
Automated internal switching of up to 4 wash buffers
|
Flow Rates
|
1 to 11, slow to fast, including low flow cell wash rates
|
Wash Cycles
|
1 to 250
|
Dispense Precision
|
≤ 3% CV: 300 µL/well
|
Residual Volume
|
≤ 2 µL/well (solid bottom plates)
|
Shaking
|
Programmable in min/sec, up to 60 minutes with slow, medium, or fast intensities
|
Soak Time
|
Programmable in min/sec, up to 60 minutes
|
Wash Speed
|
96-well: 3 cycles, 300 µL/well, 96-tube manifold, final aspirate: ≤ 30 seconds
|
User Interface
|
High resolution 5.7" touch screen with color graphical display
|
Onboard Software
|
Intel Atom processor running Windows CE
|
Dimensions (D x W x H)
|
17 x 14 x 10 in. (43.2 x 35.6 x 25.4 cm)
|
Weight
|
36 lb. (16.5 kg.)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
Click Here to Expand The 405 TS Series Model Comparison
BioTek 405 TS Series Model Comparison
Model
|
384-Well
|
96-Well
|
Cell Washing
|
4-Buffer Switching
|
Ultrasonic Advantage
|
Verify Technology
|
Biomagnetic Separation
|
Vacuum Filtration
|
405TSHT
|
x
|
|
|
|
|
|
x
|
x
|
405TSHTV
|
x
|
|
|
x
|
x
|
|
x
|
x
|
405TSU
|
x
|
x
|
x
|
|
|
|
x
|
x
|
405TSUV
|
x
|
x
|
x
|
x
|
|
|
x
|
x
|
405TSUS
|
x
|
x
|
x
|
|
x
|
|
x
|
x
|
405TSUSQ
|
x
|
x
|
x
|
|
x
|
x
|
x
|
x
|
405TSUVS
|
x
|
x
|
x
|
x
|
x
|
|
x
|
x
|
405TSUVSQ
|
x
|
x
|
x
|
x
|
x
|
x
|
x
|
x
|
405TSR
|
|
x
|
|
|
|
|
x
|
x
|
405TSRV
|
|
x
|
|
x
|
|
|
x
|
x
|
405TSRS
|
|
x
|
|
|
x
|
|
x
|
x
|
405TSRSQ
|
|
x
|
|
|
x
|
x
|
x
|
x
|
405TSRVS
|
|
x
|
|
x
|
x
|
|
x
|
x
|
405TSRVSQ
|
|
x
|
x
|
x
|
x
|
x
|
x
|
x
|
- Install new waste and buffer bottles to ensure proper fluid containment and system functionality.
- Integrate a new pump into the system, verifying compatibility and secure connections.
- Verify that the power supply and PCB output the correct voltages for reliable component operation.
- Conduct a self-test on the microplate washer, evaluating all operational parameters.
- Inspect the X-axis motor belt for signs of wear, such as tears, frays, or breakages, replacing if necessary.
- Examine dispense and aspirate tubes for any damage that could affect fluid flow accuracy.
- Confirm internal pump functionality, replacing seals as needed to maintain optimal performance.
- Inspect all internal wiring for frays, tears, or breakages, addressing any faults for electrical safety.
- Verify functionality of all internal and external sensors, ensuring accurate readings across the unit.
- Disinfect the unit using an appropriate disinfection cycle to ensure sterility and safety.
- Check that residual aspiration levels are within the manufacturer’s specified limits for accuracy.
- Conduct a final cosmetic inspection to verify that all external surfaces meet appearance standards.
Technician Inspection and Recertification Checklist: Microplate Washer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Install new waste and buffer bottles | Completed | |
Integrate new pump into system, verifying secure connections | Completed | |
Verify power supply and PCB output correct voltages | Completed | |
Conduct microplate washer self-test and evaluate results | Completed | |
Inspect X-axis motor belt for wear, replacing if necessary | Completed | |
Examine dispense and aspirate tubes for any damage | Completed | |
Confirm internal pump functionality, replace seals as needed | Completed | |
Inspect internal wiring for frays, tears, or breakages | Completed | |
Verify functionality of all internal and external sensors | Completed | |
Disinfect unit with appropriate disinfection cycle | Completed | |
Check residual aspiration levels meet specifications | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|