BioRad PowerPac Universal Electrophoresis Power Supply
The PowerPac Universal electrophoresis power supply fits the broadest range of applications of all power supplies on the market, from mini vertical and high-throughput electrophoresis to blotting. All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioRad PowerPac Universal Electrophoresis Power Supply includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
BioRad PowerPac Universal Power Supply Specifications
Output specifications
|
500 V, 2.5 A, 500 W
|
Output range (programmable)
|
10 – 500 V, fully adjustable in 1 V steps
0.01
– 2.5 A, fully adjustable in 0.001 A steps
1 – 500 W, fully adjustable in 1 W steps
|
Type of output
|
Constant voltage, constant current, or constant power with automatic
crossover
|
Timer control
|
1 min – 99 hr 59 min, fully adjustable
|
Programmable methods
|
Stores 9 methods, each with up to 9 steps
|
Display
|
128 x 64 pixel, backlit graphics LED
|
Dimensions (W x L x H)
|
27.5 x 34 x 10 cm (10.8 x 13.4 x 3.9 in)
|
PowerPac Universal Power Supply ManualClick Here to View all the Technical Specs for the Biorad PowerPac Universal
- Verify internal power supply output for correct voltage in alignment with equipment specifications
- Inspect LED display for any dim or dead pixels, ensuring optimal readability
- Examine internal wiring and connections, checking for any rips, tears, or compromised insulation
- Ensure fan is clean and fully operational to maintain adequate cooling
- Test PCB test points for proper voltage output, confirming consistent circuit performance
- Validate functionality of all buttons and switches, ensuring responsive control
- Measure and confirm constant voltage, amplitude, and wattage using precise resistance values
- Cross-check displayed voltage with multimeter readings for accuracy
- Monitor for rapid resistance changes, software errors, and power fluctuations
- Inspect line cord for any signs of wear, cuts, or damage impacting safety
- Confirm cosmetic condition meets quality standards, free of any damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Electrophoresis Power Supplies
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify internal power supply output voltage
|
Completed
|
|
Inspect LED display for dim or dead pixels
|
Completed
|
|
Examine internal wiring for rips or tears
|
Completed
|
|
Ensure fan is clean and fully operational
|
Completed
|
|
Test PCB test points for voltage accuracy
|
Completed
|
|
Validate functionality of buttons and switches
|
Completed
|
|
Confirm constant voltage, amplitude, and wattage with resistance
|
Completed
|
|
Cross-check displayed voltage with multimeter
|
Completed
|
|
Monitor for rapid resistance changes and power fluctuations
|
Completed
|
|
Inspect line cord for wear or damage
|
Completed
|
|
Confirm cosmetic condition meets quality standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|