BioRad PowerPac HV High-Voltage Power Supply
The PowerPac HV Power Supply supports an output of 5,000 V, 500 mA, and 400 W, which allows its use for all high-voltage applications, including low-current applications in the microampere range. It is ideal for IEF and DNA sequencing. With 400 W output, the PowerPac HV offers enough power to run the most demanding IEF experiments or up to four DNA sequencing cells simultaneously. All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioRad PowerPac HV High-Voltage Power Supply includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
BioRad PowerPac HV Power Supply Specifications
Output (programmable)
|
Voltage 20 – 5,000 V
Current 0.01 – 500 mA
Power 1 – 400 W
|
Type of output
|
Constant voltage, constant current, constant power, or constant
temperature (with automatic crossover)
|
Timer control
|
1 min – 99 hr 59 min, fully adjustable
|
Display functions
|
128 x 64 pixel, yellow-green backlit graphics LED
|
Programmable methods
|
Stores up to 9 basic and 9 IEF methods, each with up to 9 steps
|
Temperature control
|
30 – 90°C ±2°C
|
Safety features
|
No-load detection, sudden load change detection, ground leak
detection, overload/short circuit protection, overvoltage protection
|
Dimensions (W x L x H)
|
7.5 x 34 x 10 cm (3 x 13.4 x 3.9 in)
|
PowerPac HV Power Supply ManualClick Here to View all the Technical Specs for the Biorad PowerPac HV
- Verify internal power supply output for correct voltage in alignment with equipment specifications
- Inspect LED display for any dim or dead pixels, ensuring optimal readability
- Examine internal wiring and connections, checking for any rips, tears, or compromised insulation
- Ensure fan is clean and fully operational to maintain adequate cooling
- Test PCB test points for proper voltage output, confirming consistent circuit performance
- Validate functionality of all buttons and switches, ensuring responsive control
- Measure and confirm constant voltage, amplitude, and wattage using precise resistance values
- Cross-check displayed voltage with multimeter readings for accuracy
- Monitor for rapid resistance changes, software errors, and power fluctuations
- Inspect line cord for any signs of wear, cuts, or damage impacting safety
- Confirm cosmetic condition meets quality standards, free of any damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Electrophoresis Power Supplies
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify internal power supply output voltage
|
Completed
|
|
Inspect LED display for dim or dead pixels
|
Completed
|
|
Examine internal wiring for rips or tears
|
Completed
|
|
Ensure fan is clean and fully operational
|
Completed
|
|
Test PCB test points for voltage accuracy
|
Completed
|
|
Validate functionality of buttons and switches
|
Completed
|
|
Confirm constant voltage, amplitude, and wattage with resistance
|
Completed
|
|
Cross-check displayed voltage with multimeter
|
Completed
|
|
Monitor for rapid resistance changes and power fluctuations
|
Completed
|
|
Inspect line cord for wear or damage
|
Completed
|
|
Confirm cosmetic condition meets quality standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|