Benchmark MyTemp 65HC Digital Cooling Incubator
The Benchmark MyTemp 65HC Digital Cooling Incubator is designed to provide accurate temperature control from Ambient -25°C to 60°C with a uniformity of ± 1.0°C at 37°C. The broad temperature range of the instrument expands the use of this incubator to both heating and cooling applications. A large, digital display allows for visualization of the chamber temperature and the remaining incubation time, while an internal light is included for improved visualization of the chamber. In addition, a built-in internal outlet can be used to provide power to small instruments, such as shakers, rockers, etc. With forced air technology and stainless-steel inner construction, uniform heat transfer is ensured throughout the internal chamber. When set for sub-ambient temperatures, the powerful compressor is activated, quickly cooling the chamber to the desired temperature. With an exterior height of 39.35 inches, this incubator is designed to be floor-standing, yet is also compact enough for bench-use. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Benchmark MyTemp 65HC Digital Cooling Incubator includes:
- > 2 Adjustable Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Benchmark MyTemp 65HC Digital Cooling Incubator Specifications
Temperature Range
|
Ambient -25°C to 60°C
|
Temperature Uniformity
|
± 1.0°C at 37°C
|
Temperature Accuracy
|
0.5°C at 37°C
|
Temperature Increments
|
± 0.1°C
|
Volume
|
2.3 cu. ft. (65L)
|
Refrigerant
|
R134A
|
Internal Dimensions (W x D x H)
|
15.75 x 12.75 x 19.75 in. (40 x 32.4 x 50.2 cm)
|
External Dimensions (W x D x H)
|
21.25 x 21 x 39.35 in. (54 x 53.34 x 99.85 cm)
|
Weight
|
117 lb. (53 kg.)
|
Electrical Requirements
|
120V, 50/60 Hz, 385W, 3.2A
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|