Benchmark MyBath 12L Digital Water Bath
The Benchmark MyBath 12L Digital Water Bath is a high-precision laboratory instrument designed for superior performance and accuracy. Featuring a temperature range from Ambient +5°C to 100°C, it maintains a temperature uniformity of ± 0.2°C and accuracy of ± 0.5°C at 37°C. The digital control system allows for temperature adjustments in 0.1°C increments, displayed on a large 3-digit screen. This 12L capacity bath has the internal dimensions of 12.8 x 10.4 x 7.9 inches (32.5 x 26.4 x 20.1 cm) providing ample working space. The MyBath series stands out for its compact design, fitting various laboratory sizes. Key features include the proprietary Quik-CAL for easy in-lab calibration, high-efficiency insulation for temperature stability, and a cool-to-touch exterior. The seamless stainless-steel chamber ensures durability, ease of cleaning, and resistance to rust and contamination. The 12L model also boasts a convenient drain system for water changes, a removable see-through cover with a rear hinge, and a gabled lid design to divert condensation away from samples. This makes the Benchmark MyBath 12L Digital Water Bath an ideal choice for laboratories seeking a reliable and efficient water bath solution. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Benchmark MyBath 12L Digital Water Bath includes:
- > Bath Lid
- > Calibration
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Benchmark MyBath 12L Digital Water Bath Specifications
Temperature Range
|
Ambient +5°C to 100°C
|
Temperature Uniformity
|
± 0.2°C at 37°C
|
Temperature Accuracy
|
± 0.5°C at 37°C
|
Control
|
Digital
|
Capacity
|
12L (3.2 gal.)
|
Bath Dimensions (W x D x H)
|
12.8 x 10.4 x 7.9 in. (32.5 x 26.4 x 20.1 cm)
|
Overall Dimensions (W x D x H)
|
18 x 10.5 x 16.2 in. (45.7 x 26.7 x 41.1 cm)
|
Weight
|
17 lb. (7.7 kg)
|
Electrical Requirements
|
100-120V, 60Hz, 900W
|
- Confirm power supply and PCB are outputting stable, correct voltages across all essential circuits to ensure reliable operation
- Inspect the display for any dim, flickering, or dead pixels, replacing any compromised sections to maintain full clarity and accurate readouts
- Verify that all buttons, knobs, and switches are functioning properly, providing precise and responsive control across all settings
- Conduct a thorough inspection of the water reservoir for any holes, rust, or other structural damage that may impact its integrity
- Fill the reservoir with water, checking closely for any external or internal leaks that could affect performance
- Examine the lid for any signs of wear or damage, replacing hinges and handles if necessary to ensure a secure and proper seal
- Inspect for any internal leaks originating from the water reservoir, confirming the unit’s internal plumbing integrity
- Conduct a comprehensive inspection of internal wiring for any breakages, tears, or wear; replace any underperforming or damaged components to maintain safe operation
- Examine internal insulation for signs of burns, breaks, or deterioration, replacing as needed to ensure thermal efficiency and safety
- Confirm that all feet are intact and that the bath sits level and balanced on a flat surface to prevent movement or instability
- Calibrate both temperature accuracy and uniformity using a NIST-traceable temperature probe, verifying compliance with specification standards
- Perform a detailed cosmetic inspection to confirm that all external surfaces and finishes meet aesthetic and quality standards
Technician Inspection and Recertification Checklist: Water Bath
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm power supply and PCB output stable, correct voltages across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace for clarity
|
Completed
|
|
Verify functionality of all buttons, knobs, and switches for responsive control
|
Completed
|
|
Inspect water reservoir for holes, rust, or structural damage
|
Completed
|
|
Fill reservoir with water; check for external or internal leaks
|
Completed
|
|
Examine lid for wear or damage; replace hinges or handles as needed
|
Completed
|
|
Inspect for internal leaks from water reservoir to confirm plumbing integrity
|
Completed
|
|
Check internal wiring for damage; replace underperforming components for safety
|
Completed
|
|
Examine internal insulation for burns, breaks, or deterioration; replace as needed
|
Completed
|
|
Confirm feet are intact and bath is balanced on flat surface
|
Completed
|
|
Calibrate temperature accuracy and uniformity with NIST-traceable probe
|
Completed
|
|
Perform cosmetic inspection to confirm quality of external surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|