Benchmark MultiTherm Heating and Cooling Shaker
Speed Range 200 to 1,500 RPM
Temperature Range Ambient –20C to 100C
Maximum Capacity 6 x 50ml
- Temperature control from Ambient -20 to 100C
- Exchangeable blocks, for tube sizes 0.2 to 50 ml
- Precision contoured wells for uniform thermal transfer
- Sequence link function for linking or repeating programs
- Compatible with a variety of applications including denaturation of DNA, RNA and proteins, as well as lipid extractions
- Compact 8 x 11 in. footprint for efficient use of laboratory space
The Benchmark MultiTherm Heating and Cooling Shaker is a temperature controlled vortexer that features a temperature range of Ambient -20°C to 100°C and a speed range of 200 to 1,500 RPM. Despite its modest 8 x 11 inch footprint, all popular tubes and plates are compatible when in use with the appropriate block, including as many as six 50ml tubes. Speed, time, and temperature settings are continuously visible on the LCD display screen, simultaneously showing both actual and selected values. Integral over-temperature control is featured and ensures long life, safety, and sample integrity. The Benchmark MultiTherm Heating and Cooling Shaker is useful for a variety of molecular biology applications including denaturation of DNA, RNA, and proteins, lipid extractions, yeast and bacteria cultivation, restriction digests, reverse transcription, and more.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Benchmark MultiTherm Heating and Cooling Shaker includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Benchmark MultiTherm Heating and Cooling Shaker Specifications
Speed Range
|
200 to 1,500 RPM
|
Speed Increments
|
± 10 RPM
|
Temperature Range
|
Ambient -20C to 100C
|
Temperature Accuracy
|
± 1.0C
|
Temperature Increments
|
± 0.1C
|
Heating Rate
|
Heating 5°C/minute
|
Timer
|
1 minute to 99 hours 99 minutes, or continuous
|
Orbit
|
2mm (0.078 in.)
|
Dimensions (W x D x H)
|
8.25 x 11.5 x 8.25 in. (21 x 29 x 21 cm)
|
Weight
|
18 lb. (8 kg.)
|
Electrical Requirements
|
115V, 60Hz, 4A, 200W
|
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|