Beckman Coulter X-12 Centrifuge
Maximum Speed
Fixed Angle Rotor: 10,200 RPM
Swinging Bucket Rotor: 4,750 RPM
Maximum Capacity
4 x 750 ml
- Offers 10 user-defined programs for customizable operation and flexibility in lab workflows
- Features 10 acceleration and 11 deceleration profiles for precise sample handling
- Auto rotor recognition ensures safe and accurate speed settings
- Toggle between RPM and RCF display modes for user convenience
- Ideal for diverse applications, including cell isolation, blood separation, and protein precipitation
- Quiet operation with a noise level of ≤ 68 dBA for a comfortable lab environment
- Durable design with compact dimensions of 34.3 x 76.2 x 62.2 cm (13.5 x 30 x 24.5 in)
- Operates on a 208V, 60Hz power supply
The Beckman Coulter X-12 Centrifuge is ideal for simplifying separation processes and saving time whether harvesting cells or separating blood components. The X-12 offers speeds up to 10,200 RPM with a maximum RCF of 11,400 x g. Featuring 10 acceleration and 11 deceleration profiles, as well as 10 user-defined programs, time options, toggling between display modes, RPM and RCF set, and auto rotor recognition, the Allegra X-12 offers robust flexibility for a variety of uses in the lab.
This rotor reaches a maximum speed of 3,750 RPM.
Applications include: routine processing, such as sample preparations, concentrations, receptor binding, extractions, pelleting, phase separations, purifications, and column centrifugations; cell isolation; biding studies and separation of whole blood; processing large numbers of small-volume samples in multiwell plates for concentrating tissue-culture cells, receptor binding, in-vitro cytotoxicity studies, cloning and replicate studies, and genetic engineering experimentation; and rapid sedimentation of protein precipitates, cell debris, and large particles.
All of our equipment is fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Beckman Allegra X-12 Centrifuge includes:
- > SX4750 Swing Bucket Rotor
- > Set of 4 x SX4750 Round Buckets
- > Set of 2 x 15ml Adapters
- > Set of 2 x 50ml Adapters
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Beckman Coulter X-12 Centrifuge Specifications
Maximum Speed
|
Fixed angle rotor: 10,200 RPM
Swinging bucket rotor: 4,750 RPM
|
Maximum g-Force
|
11,400 x g
|
Maximum Capacity
|
4 x 750 ml
|
Set Speed
|
actual rotor speed ±
10 rpm or equivalent RCF
|
Temperature
|
Factory set at 20°C
|
Acceleration
|
10 acceleration rates
|
Deceleration
|
11 deceleration rates
|
Set Time
|
up to 9 hr 59 min or
continuous
|
Humidity Restrictions
|
< 75%
(noncondensing)
|
Noise Level
|
≤ 68 dBA
|
Dimensions (H x W x D)
|
34.3 x 76.2 x 62.2 cm
(13.5 x 30 x 24.5 in)
|
Electrical
Requirements
|
208V, 60Hz
|
Refrigeration
|
No Refrigeration
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
Beckman Allegra X-12 Manual
Click Here to View the Technical Specs for the Allegra X-12
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|